19
Note: Choice will determine whether
the units are displayed in mS
or PPT. COND = mS or TDS
= PPT.
2.
Press
repeatedly until
“SALt OFF” is displayed
(see Figure 12).
3.
Press
. The display will
change from SALt OFF to
“SALt On” (see Figure 13).
4.
Press
. All COND or
TDS readings, as selected, are now displayed in Salinity units.
Either COND/mS units or TDS/PPT units are now selected.
If you wish BOTH COND and TDS to be displayed in Salinity
units, repeat with the other measurement key.
XIII. TOTAL RETURN to FACTORy SETTINGS “FAC SEL”
There may come a time when it would be desirable to quickly reset all the
recorded calibration values in the instrument back to the factory settings.
This might be to ensure all calibrations are set to a known value, or to
give the instrument to someone else free of adjustments for a particular
application. NOTE: All stored data will be lost.
1.
Press any measurement key.
2.
Press
repeatedly until
“FAC SEL” is displayed
(see Figure 14).
3.
Press to accept the resetting. Display will return to COND.
XIV.
CARE and MAINTENANCE
Your TechPro
II
should be rinsed with clean water after each use.
Figure 12
442
TDS
Figure 13
442
TDS
PPT
Figure 14
Содержание TechPro II
Страница 1: ...18 February 16 TECHPRO Operation Manual MODELS TPH1 TP1...
Страница 2: ......
Страница 4: ...ii...
Страница 38: ...34 XXIII ADDENDUM...
Страница 39: ...35...
Страница 40: ...36 XXIV NOTES...
Страница 41: ...37...
Страница 42: ...38...
Страница 43: ......