PoolPro PS9 Quick Start Guide
Handheld Conductivity, Mineral / Salt Concentration, TDS, ORP, Hardness, Alkalinity, LSI, Free Chlorine, pH, and Temperature Meter
PS9QSI REV: 07-16
This Quick Start Guide is for the Myron L
®
Company Pool Pro Model PS9. Download the full PS9 Operation Manual (P/N PS9OM) from www.myronl.com to get more
detailed instructions for taking measurements (including Alkalinity Hardness and LSI calculation), storing or recalling measurements, changing instrument settings,
maintenance procedures, calibration procedures, Troubleshooting Guide and FACTORY CAL reset.
QUICK REFERENCE INSTRUCTIONS
Press and release any one of these buttons to turn on the PS9 and begin taking measurements.
Press to calculate LSI.
Press and release the CAL/MCLR button to calibrate (CAL) the active measurement parameter or press and hold to clear displayed memory location (MCLR).
Press and release to store (MS) a measurement or move upward on the list of stored measurements (Location 1 to 100) while in memory recall mode.
Press and release to view the most recent stored measurements or move downward on the list of stored measurements (Location 100 to 1). Also used to enter
Settings Mode.
PS9
FEATURES
1.
TEMPERATURE SENSOR –
Houses thermistor for making temperature measurements.
2. CONDUCTIVITY CELL
–
Contains electrodes that generate a flux field in defined area for all measurements except
pH, ORP, and FC
E
.
3. PH SENSOR
(user replaceable) –
Measures test sample’s hydrogen ion concentration
. The reference junction is
located under the glass bulb.
4.
pH SENSOR PROTECTIVE CAP
– Snaps in place to protect pH Sensor when not in use and seals in Storage
Solution.
5.
LCD DISPLAY –
Displays measurements, units of measure, active measurement mode, current solution mode
setting, low battery warning (LOBATT), and other information (see below).
6. MEASUREMENT KEYS
– Press and release any one of these to turn on the PS9
and begin taking measurements.
7. FUNCTION KEYS
–
Press and release to store the current measurement (MS), recall a stored measurement (MR).
Also used to move between memory locations and to select various mode settings
.
8. CAL/MEMORY CLEAR KEY
– Use to calibrate (CAL) the active measurement parameter, enter or clear the
contents of the currently displayed memory location (MCLR).
9
.
BLUDOCK SYMBOL
– Indicates that this PoolPro PS9 is equipped with Myron L
®
Company’s optional wireless
transceiver for wireless data download.
10
.
CELL EXTENDER –
Holds sample and reagents during Alkalinity and Hardness measurements.
11. PLUNGER
– Used to create proper cell cup volume of liquid during Alkalinity and Hardness measurements.
12
.
CELL EXTENDER CAP
– Seals cell extender while mixing sample and reagents during Alkalinity and Hardness
measurements.
LCD DISPLAY
1. MAIN VALUE DISPLAY
– Displays the current measurement value and settings menu information. During Alkalinity & Hardness titrations, prompts appear here at
various stages of the process.
2. UNITS OF MEASURE ICONS
– Appropriate icon appears to show the units of measure for the current measurement.
3.
TEMPERATURE UNITS OF MEASURE
– Displays °C (Celsius) or °F (Fahrenheit) based on setting selected for
Temperature readings.
4.
MEASUREMENT ICONS
– Appropriate icon appears to show what type of measurement is being made.
5.
BUFFER ICON
– Appears when the PS9 is in pH Calibration mode to indicate that the instrument is expecting a pH
buffer solution.
6.
TEMPERATURE DISPLAY
– Displays the temperature of the sample solution, memory location number
(1 to 100) or buffer during pH calibration.
7.
SOLUTION MODE ICONS
– Appropriate icon appears to indicate the current solution temperature compensation
mode setting for Conductivity, Min/Salt and TDS measurements.
8.
LOBATT ICON
– Appears when the PS9
’s battery requires replacement.
9. CAL ICON –
Appears when the PS9
is in Calibration mode.
10.
MEMORY ICON
– Appears to indicate that the values and icons being displayed are for a measurement stored in the PS9
’s memory (NOT a live measurement).
OPERATING INSTRUCTIONS
DEFAULT SETTINGS:
•
Temperature Units:
°C (Temperature is displayed in Degrees Celsius)
•
Solution Mode: Conductivity – KCl
MEASUREMENT SETUP:
Before you take a reading, make sure the PS9
is
clean, calibrated, and if measuring COND or TDS, that the desired
solution mode has been selected (See Section II of Operating
Instructions, below). The sample solution must also be within the
specified range.
I. MEASUREMENT
1.
Rinse and empty the appropriate sensor (conductivity or pH / ORP) 3 times
with sample to be measured.
2.
Fill with test sample.
3.
Press appropriate Measurement Key:
COND to measure CONDUCTIVITY
0-9999 MICROSIEMENS (µS/cm), 10-200 (mS)/cm
TDS: Measures TOTAL DISSOLVED SOLIDS
0-9999 parts per million (ppm), 10-200 (ppt)
MIN / SALT: Same as TDS but displays in
0 - 200 parts per thousand (ppt).
displays as PPT of NaCl
ORP: Measures Oxidation Reduction Potential (REDOX) ± 999 mV
pH: Measures 0-14 pH
Press one of these to preform an Alkalinity or Hardness
measurement.
Press to calculate LSI. Uses stored or hypothetical Alkalinity and
Hardness data.
•
For complete instructions on performing these measurements,
download the full PoolPro PS9 Operation Manual and watch our
YouTube videos (
Search for: Ultrameter III 9P and 9PTK
).