Maintenance

batteRy check:

Model  D-1  has  a  battery  indicator  glow  light  visible  

through  the  small  hole  on  the  lower  right-hand  corner  

of  the  meter  face  plate.  if  this  light  fails  to  glow  when  the  

black button is pressed, replace both batteries.

batteRy ReplaceMent:

Detach  the  battery  connectors.  pull  on  the  plastic  

straps  to  remove  the  batteries.  replace  with  fresh  zinc  

carbon  or  alkaline  9  volt  batteries.  reinsert  the  plastic  

straps to secure batteries. replace the bottom cover.

cell cup:

self-conditioning  of  the  built-in  electrodes  occurs  

each  time  the  button  is  pressed  with  a  sample  in  the  cell 

cup.  this  ensures  consistent  results  each  time.  With  

some  samples  a  small  downward  swing  of  the  pointer  is  

a  result  of  this  conditioning  action.  this  action  is  pow- 

erful  and  removes  normal  films  of  oil  and  dirt.  However, 

 

if  samples  are  allowed  to  dry  in  the  cell  cup,  a  film  will 

 

build  up.  This  film  reduces  accuracy.  When  there  are 

 

visible  films  of  oil,  dirt,  or  scale  in  the  cell  cup  or  on  the

 

electrode,  scrub  them  lightly  with  a  small  brush  and  

household  cleanser.  thoroughly  rinse  out  the  cleanser,  

and the meter is ready for accurate measurements.

SolutionS inSiDe the MeteR

Your  Dialysate  Meter  is  a  rugged  instrument  and  will 

withstand  dialysate  exposure  around  its  cell,  meter  

movement,  and  switches.  However,  care  should  be  

taken  to  keep  solutions  from  leaking  in  around  the  

bottom  cover.  it  is  not  sealed  (to  prevent  condensation  

from forming).

Large  amounts  of  calibration  solution  or  dialysate  will 

almost  certainly  damage  the  meter  movement  or  

electronics.  such  meters  should  be  returned  to  the  

Myron L company for repair.

if  there  are  only  a  few  drops  inside  the  meter,  dry  it  as 

follows:

1.  shake  excess  solution  out  of  the  inside  of  the  meter.

2.  Dab  the  exposed  surfaces  dry  with  an  absorbent  

  cloth  or  tissue.  avoid  pushing  any  liquid  into  the  

  calibration control or push button switch.

3.  air  dry  the  meter  in  a  warm  area  with  the  bottom  

  cover  off.  allow  several  hours  for  thorough  drying.

if  the  dialysate  entered  through  a  leak  in  the  case  or  

cell,  or  if  the  instrument  shows  erratic  readings  or  other  

unusual  behavior,  return  it  to  the  Myron L  company  for 

service.

SeRvice

  if your D-1 meter fails to operate properly:

1.  check  the  glow  light  to  make  sure  the  batteries  are  

  good.

2.  check  the  cell  cup  and  electrodes  to  ensure  they  are 

  clean.  any  contamination  could  alter  the  reading.  see 

  Maintenance.

any  service  required  other  than  battery  replacement, 

cleaning,  or  calibration  must  be  referred  to  the  Myron L 

company.  When  returning  an  instrument,  please  pack  

it  securely  and  enclose  a  note  specifying  the  problem.  

the  plunger  should  be  included,  but  Do  not  ship  with  

plunger in cell (damage could result).

oRDeRing

 

to  order  accessories  or  instruments,  contact  your  

nearest  stocking  distributor,  or  the  Myron L  company.

aDDitional pRoDuctS:

to  see  the  full  range  of  Myron L  products,  go  to  our  

web site @ www.myronl.com.

the Myron L Dialysate Meters have a limited two year war- 

ranty.  if  your  instrument  fails  to  operate  properly,  check  the 

batteries  and  calibration.  if  it  still  fails  to  function  properly,  

return it prepaid to the Myron L company.

Faulty  instruments  may  be  returned  to  us  without  prior  

authorization.

MeteRS within two yeaR waRRanty peRioD:

Failures  due  to  materials  or  workmanship  will  be  repaired  

or replaced (our option) without charge if returned freight pre- 

paid. if failure is deemed by the factory to have been caused  

by abuse or tampering, the following procedure will apply.

inStRuMentS/contRolS out of waRRanty:

Diagnosis  will  be  made  and  repairs  completed,  providing  

the repair charges are $70.00 or less.

note:

 actual repair charges may be less than this amount.

We  will  diagnose  (but  not  repair)  a  returned  meter  and  

send an estimate of charges if anY of the following apply:

1.  repair charges will be more than $70.00.

2.  You specifically request an estimate of required repairs and 

 

 

charges.

3.  the cost of required repairs exceeds one-half the list price  

 

of a new instrument.

4.  the  instrument  is  over  ten  years  old.  because  of  compo- 

 

nent changes and improvements, such instruments can no 

 

longer be repaired.

note:

  unrepaired  meters  are  discarded  unless  you  want  

them  returned  to  you.  if  so,  there  is  currently  a  $35.00  charge  

per unrepaired instrument to cover diagnosis and handling.

this  warranty  is  limited  to  the  repair  or  replacement  of  the  

Myron L  Dialysate  Meter  only.  the  Myron L  company  

assumes no other responsibility or liability.

waRRanty/SeRvice

 

Specifications subject to change without notice

2450 impala Drive, carlsbad, ca 92010-7226

teLepHone: +1-760-438-2021

Fax: +1-760-931-9189

e-Mail: [email protected], [email protected]

www.myronl.com

Отзывы: