Single Range Dialysate Meter™

operating instructions

User’s Manual for Model D -1

this  manual  describes  your  Myron L  D-1  meter,  tells  

you  how  to  use  it,  and  how  to  keep  it  working  accurately  

for many years.

Your  Model  D-1  single  range  Dialysate  Meter™  is  a 

compact  instrument  which  operates  on  the  principle  of 

electrical  conductivity.  it  will  quickly  determine  if  dialy- 

sate  concentrations  are  correct.  its  dial  provides  an  

expanded  11-15  millimhos  (millisiemens)  scale  for  ac- 

curate readings.

a  unique  feature  of  Model  D-1  is  the  green  color  band 

above  the  dial.  this  provides  an  immediate  visual  con- 

firmation  of  whether  or  not  the  dialysate  concentration 

 

is  within  the  patient’s  “safe  zone”.  the  color  band  is  

adjustable  and  can  be  pre-set  by  the  doctor  or  techni- 

cian  for  each  patient.  adjustment  is  made  by  turning  

the  screw  on  the  face  of  the  meter.  this  screw  may  be  

covered with tape or a label to avoid tampering.

the  D-1  is  3.4”  x  4.5”  x  4.0”  (85  x  129  x  126mm)  and  

weighs  less  than  one  pound  (0.45kg).  D-1  meters  are 

completely  self-contained.  the  built-in  cell  is  automat- 

ically  temperature  compensated  from  60º  to  110ºF  

(16º  to  43ºc).  they  are  powered  by  two  9  volt  batteries  

good  for  a  least  2000  tests,  or  one  year  shelf  life.  a  light  

bulb,  which  glows  when  the  black  button  is  pushed,  

serves as a battery reserve indicator.

the  pictures  in  this  manual  show  the  major  operating  

parts  of  your  Myron L  D-1  Dialysate  Meter.  Handle  your  

D-1 and identify these parts to become familiar with it.

Battery

Indicator

Plunger

Cell Cup

Pointer

Button

Upper

Electrode

Color Band

Adjust

DeScRiption