Myron L 758II-121-SC Скачать руководство пользователя страница 4

connected to an appropriate Sensor, and that the Sensor is 
immersed in water within the range that the Monitor/controller will 
be required to read; and the front panel is removed. 
 

1.   Make a note of the reading on the display.

 

2.   While pressing the Calibration/Full Scale Test Switch 

 

 

(FSSW), verify that the front panel display is indicating 

 

 

a full scale reading. If not, see Calibration, section V.C.

 

3.   Press and hold the “SET POINT” switch on the front 

 

 

panel. Using a tweaker or a small screwdriver, adjust  

 

 

the Set Point trimmer adjustment screw on the circuit 

 

 

board to sweep the display from zero to full scale. (A  

 

 

digital display may be blank at the full scale end. This  

 

 

is normal.) Listen for the alarm relay to click on and off  

 

 

as the alarm set point moves past the water reading.

 

4.  Adjust the alarm to the desired set point value. Release  

 

 

the “SET POINT” switch. 

NOTE:

 For Models with SC/SCO module, repeat STEPS 3 & 4 to 

check out Set Point #2. 

QUICK SET POINT ADJUSTMENT - 

 

 

 

 

 

 

 

See Section IV.C.3.

The set point setting is based upon the user’s particular water 
purity specifications or requirements. 

NOTE:

 The optional 

second

 relay/alarm is “stacked” on the first relay/alarm, 

therefore, when setting the optional 

second

 relay/alarm Set 

Point, the #1 Set Point must be ‘set’ first. 
 

1.   While pressing the “SET POINT” switch, turn the Set    

 

 

Point #1 adjustment screw (see figure V.A.1) until the   

 

 

desired set point value is indicated on the display.

HYSTERESIS (DEAD BAND) ADJUSTMENT - 
 

 

 

 

 

 

 

See Section IV.C.4.

PRIMARY COMPONENT IDENTIFICATION - 
 

 

 

 

 

 

 

Section V.A.

Review the figure below to familiarize yourself with the Main 
circuit board assembly. The diagram has the second alarm/
control module option installed. 

QUICK CALIBRATION - Section V.C.

WARNING:  When performing calibration procedures,
the technician must take extreme care to avoid
contacting the circuitry other than the CALibration
control. Failure to do so could result in damage to
the equipment, property and/or personal injury. 

The following assumes the front panel has been
removed and the power is ON.

 

ELECTRONIC CALIBRATION (CIRCUIT ONLY) - 
 

 

 

 

 

 

 

See Section V.C.1.

Full Scale Adjustment

 

1.  Press and hold the Full Scale Test switch. The display 

 

 

should indicate Full Scale for the particular range  

 

 

 

selected, i.e. 0-500 ppm should indicate 500. If not, set  

 

 

to Full Scale with the 

CAL

ibration control.

 

2.  Turn power 

OFF

.

 

3.  Re-install front panel as described in “REASSEMBLY”.

 

4.  To operate, turn power 

ON

.  

10VDC Calibration - See Section V.C.1.b.

Using Standard Solutions - Section V.C.2.

The

 BEST

 method of verifying and recalibrating your 

conductivity/TDS Monitor/controller is with NIST traceable 
Standard Solution (available from the Myron L Company). 
Because it includes the sensor, the entire system is recalibrated.  

NOTE:

 Since standard solution calibrations are NOT practicable 

with resistivity models, another means of verification or 
calibration of resistivity models is to use the transfer standard 
method, using a hand-held or portable instrument capable of 
resistivity measurements, i.e. the Myron L Ultrameter 

II

™. See 

section V.C.4 for description.
The following procedure describes the easiest method for 
standard solution calibration of your Conductivity/TDS
Monitor/controller. 

1.   Using a standard solution which is 60-90% of full scale  
 

of the instrument, rinse thoroughly and fill a clean glass  

 

beaker with the standard solution. 

2.  Place sensor in the beaker of standard solution. The  
 

level of standard solution must be high enough to cover

 

at least 1/2” above cross hole. 

3.  Slowly shake the sensor to remove air bubbles from    
 

inside the sensor bore hole. 

4.   Allow 5-10 minutes for temperature to equilibrate. For 
 

the quickest and the best results, both the sensor and  

 

solution should be at the same temperature. 

5.  Turn power 

ON

.

6.  If the reading is different from the standard solution,
 

adjust 

CAL

ibration control on the main circuit board    

 

until the reading matches the solution value. 

7.  After adjustment, turn power 

OFF

.

8.  Re-install front panel as described in “REASSEMBLY”.
9.  To operate, turn power 

ON

.  

SENSOR SUBSTITUTE CALIBRATION - 
 

 

 

 

 

 

 

See Section V.C.3.

TRANSFER STANDARD METHOD - See Section V.C.4.

B
L
K

W
H
T

SENSOR

RANGE

MODULE

MAIN

CALIBRATION

CONTROL

SC/SCO

OPTIONAL

SECOND

ALARM/

CONTROL

MODULE

#2 RELAY
LIGHTS &

SWITCH

CONNECTOR

RUBBER TAPE

(DO NOT

REMOVE)

RANGE

MODULE

ALIGNMENT

3SO/3SE

&/OR

3RO/3RE

SWITCH

OR

FLOW SWITCH

CONNECTION

POWER

AC LINE/ +DC

AC NEUTRAL/ -DC

GROUND

RELAY #1

}

COM

NC

NO

DISPLAY/METER
CONNECTOR

4-20 CONNECTOR

SET POINT  #1
CONVERSION

SET POINT #2 CONTROL

SET POINT  #2
CONVERSION

DISPLAY CALIBRATION
CONTROL (FACTORY SET)

SET POINT #1
HYSTERESIS
RIGHT INC
LEFT DEC

CHASSIS GROUND - OEM

SET POINT #2 HYSTERESIS
RIGHT INC / LEFT DEC

SOLID STATE (24VDC 30mW)
      OUTPUT
PA™ PIEZO ELECTRIC ALARM
      OR
RA™ REMOTE ALARM OR
CUSTOMER CONNECTION

SOLID STATE (24VDC 30mW)
      OUTPUT

RA™ REMOTE ALARM AND

PA™ PIEZO ELECTRIC ALARM
      OR
CUSTOMER CONNECTION

RELAY #2

}

COM

NO

NC

FULL SCALE
PUSH TO TEST

SET POINT
#1 CONTROL

0-10VDC OUTPUT

N
E
U

(+)

(-)

R
E
D

G
R
N

3S

PA

HYS1

SP1

FS SW

DIS

INC

DEC

CAL

INC

-121

2000µS

}

SPC

TRANSFORMER

FUSE*

115/
      230

751    756
752    757
753    758
754    759

CHS

GND

PWR

  C    GD   NC   NO   CM

C

M

  N

O

  N

C

SPC

INC

UP

+

+

PA

RA

{

BK  WT RD  GN  NU  R-  R+

REMOVE T

O INST

ALL

SECOND RELA

Y

 

5/21/03

Figure V.A.1

Содержание 758II-121-SC

Страница 1: ...els 756II 757II 758II 759II Resistivity Models 751II 752II 753II 754II 750 Series II CONDUCTIVITY TDS RESISTIVITY MONITOR CONTROLLER Operation Manual 22 October 08 ACCURACY RELIABILITY SIMPLICITY Water Quality Instrumentation Accuracy Reliability Simplicity ...

Страница 2: ...750 Series II Model 758 121 SC A Digital Conductivity Monitor controller with a Range of 0 2000 µS and a Second Alarm control Displaying Full Scale II HIGH LOW SET POINT MICROSIEMENS CM 750II ...

Страница 3: ...ll more than about 8 inches 20CM or you could damage the wiring harness REASSEMBLY 1 Carefully reinstall the front panel bottom first Ensure no wires have been pinched between enclosure and front panel 2 Reinstall the four 4 screws and tighten 3 To operate turn power ON INTRODUCTION Section I This section covers the specifications of your new Monitor controller including sensor information INSTALL...

Страница 4: ...BEST method of verifying and recalibrating your conductivity TDS Monitor controller is with NIST traceable Standard Solution available from the Myron L Company Because it includes the sensor the entire system is recalibrated NOTE Since standard solution calibrations are NOT practicable with resistivity models another means of verification or calibration of resistivity models is to use the transfer...

Страница 5: ...on L Company assumes no other responsibility or liability ADDITIONAL INFORMATION Custom Monitors controllers available contact us with your special needs Price and delivery upon request ALL SPECIAL ORDER ITEMS ARE NON RETURNABLE AND NON REFUNDABLE Because of our policy of continuous product improvement the Myron L Company reserves the right to make changes in design specifications and prices witho...

Отзывы: