DESCRIPTION

This manual describes your Myron L DS meter, tells you how to

use it, and how to keep it working accurately for many years.

Your model 512 single range DS Meter is a compact instrument

which operates on the principal of electrical conductivity. It will
quickly determine the conductivity or ppm/Total Dissolved Solids
of almost any solution. Models 512M2, 512M3, 512M4, 512M5
and 512M10 have dials calibrated in micromhos (

µ

mhos), which

is equivalent to the metric microsiemens (

µ

S). Models 512T2,

512T3, 512T4, 512T5 and 512T10 convert the conductivity directly
into parts per million (ppm) of Total Dissolved Solids.

All are 3.4" x 4.5" x 4.0" (85 x 129 x 126mm) and weigh less

than one pound (0.45kg). DS Meters are completely self-contained.
The built-in cell is automatically temperature compensated from
50

°

 to 160

°

F (10

°

 to 71

°

C). They are powered by a 9 volt battery

good for at least 2000 tests, or one year shelf life.

512T

512M

MODEL

RANGE

MODEL

RANGE

512T2

0-50 ppm

512M2

0-50 

µ

mho

512T3

0-500 ppm

512M3

0-500 

µ

mho

512T4

0-2500 ppm

512M4

0-2500 

µ

mho

512T5

0-5000 ppm

512M5

0-5000 

µ

mho

512T10

0-1O,000 ppm

512M10

0-10,000 

µ

mho

By using an RE-10 Range Extender (see ACCESSORIES), the

maximum range of most meters may be increased ten times.

The pictures in this manual show the major operating parts of

your Myron L DS Meter. Handle your DS Meter and identify these
parts to become familiar with it.

Model 512M5 shown