12. Filter cotton must be used in the dermabrasion treatment in case dirt is absorbed into the
tube and causes damages to machine.
13. 40K function can not be used on the areas next to heart, like chest, back and head.
14. Make sure the target area is kept moistened to make treatment receiver feel comfortable.
Do on perform treatments on dry skin(except for the lipo laser treatment)
15. Please inform the treatment receiver of possible sensations of micro electric currents before
treatment.
16. Cleaning the dermabration tip is a must, which can prevent cross infection. Brush dirt and
grime out of the tip, rinse the tip with clean water, soak the tip in ethanol for 3 minutes and
then rinse the tip again to remove ethanol.
17. Please make sure you can distinguish the suction hole from the air outlet. Do not connect
the oxygenation sprayer to the suction hole, otherwise the machine will be broken.
18. The liquid used in the oxygenation sprayer should be watery fluid. The possibility is there will
be blockage and machine may get broken if you pour thick skincare products into it.
19. When you use the oxygenation sprayer on the customer’s philtrum, please let customer
change her/his way of breathing.
20. Clean the instrument with normal saline after operation to ensure its cleanliness and
hygiene and prolong its service life.
21. Use machine or train machine operators in strict accordance with instructions in the user
manual.
4. Troubleshooting & Solutions
1. The machine cannot be started?
A. Make sure the power cord is connected to the socket with normal power supply.
B. Please check whether the power cord is properly connected to the socket at the back of the
machine
2. The machine can be started but the cold hammer doesn’t work?
A: The cooling fan inside is broken and then doesn’t work
B: The circuit board inside is loose or damaged
C: If aforementioned problems occur, you need to replace the handpiece.
3. The machine can be started, but there is no electric currents during the BIO treatment.
A: Please check whether the start button on the control panel is pressed.
B: Please check whether the handpiece is connected to the machine body properly.
C: Please check whether there is liquid medium on face, for there are obvious sensations only
when mediums are applied to the target area.
D: If it is not any case mentioned above, the handpiece may be broken. You need to contact
after-sale service staff for solution.
4. There is no spray coming out when using the sprayer?
A: Please check whether the nozzle of the sprayer is blocked.
B: Please check whether the sprayer is connected to the machine properly.
C: Please check whether the small lever on the sprayer comes loose.
5. There is no suction when the dermabrasion treatment is performed?
A. Please check whether the hose is connected to the machine properly.
B. Please check whether the hose and dermabrasion handpiece are connected or whether the
connection is correct.