- 13 -
“
Vm
”
in
“
A
”
field indicates actual measurement voltage monitored by the meter, which shows 1V is applied to
the sample.
“
B
”
field shows a capacitance of a 10
F product, which is about 9.86
F which is within the
specified tolerance.
Fig.19 Measure Setup Screen Showing ALC OFF
Note 2: Check Method of Measurement Voltage Applied to Capacitor Under Test
One method to check the measurement voltage actually applied to a capacitor under test is
to apply plus and minus probes of the tester to both terminals respectively of the capacitor set
into measurement jig, and read the indicated voltage.
A photograph in which
“
measurement voltage
”
of Fig.17 above is checked is shown in
Fig.10. It shows it indicates a value of 998mVrms.
Also, a photograph in which the
“
measurement voltage
”
of Fig.19 is checked is shown in
Fig.21. It shows the tester indicating 187mVrms and low capacitance due to the AC
measurement voltage being less than 1Vrms.
Fig.20 Tester Measurement Example for ALC ON State
Fig.21 Tester Measurement Example for ALC OFF State
Tester
Teste