L60iF Issue 9 November 2018
12
6.1 Basic re-calibration
Before starting the calibration procedure it is sensible to
check
the calibration against the reference. If no
significant difference is shown then the procedure is unnecessary.
As explained earlier, the minimum requirement for recalibration is a flowmeter of known accuracy and a
means of adjusting the airflow. These items can be obtained separately and are also fitted to the Munro
Instruments Pump Test Stand.
Calibration set up
NOTE If temperature calibration is also being carried out, please refer to paragraph 5.4.2. Note that
temperature calibration compensates for changing pump performance during the warm-up period to
give maximum accuracy
.
6.1.1. A flow gauge with a maximum flow of 100 litre/minute is recommended. Also required is a cone
adapter, which temporarily replaces the normal 60mm filter holder. Unscrew the filter holder and locate the
cone adapter in its place using its locking ring.
6.1.2. Connect the flow gauge to the L60iF via the Restriction Valve (see diagram above). Open the valve to
offer minimum restriction.