background image

1.

 

Overview

The MP770291 clamp is a current probe for oscilloscope which uses a Hall effect cell for the 

measurement of DC or AC current without modification of the installation (without switching off 

the circuit) 
It can measure currents from 50 mA to 100 A peak. 
It has 2 ranges and 2 lights indication: 

-“ON”, correct power supply to the clamp. 

-“OL”, overload of the range in use (saturation or peak). 

In addition, a thumbwheel can be used to reset zero for adaptation to the measurement 

environment. 
This clamp adapts to any measurement instruments which have a BNC input and an 

impedance of 1MΩ,

100pF.   

2. Specification

Range 

100mV/A 

10mV/A 

Current range 

50mA-10Apeak 

10A-100Apeak 

Error

Accuracy

 

2%±2mV 

Bandwidth

-3dB

 

DC~100kHz

 

Phase shift 

DC~65Hz

<1.5°

 

DC~65Hz

<1°

 

Load 
Impedance 

≥1MΩ

≤100pF

 

Insert 
impedance 

0

01Ω

 

Noise 

3mV 

48

0μV

 

Slew Rat 

0

.

3V/μ

20mV/μs

 

Rise/fall time 

3μs

 

4μs

 

Measurement 
Conditions

 

23 

 

±

  5 

  ,20~75%RH, Sinusoidal 

frequency :48~65Hz, External magnetic field 
<40A/m

no load flo

w,  the  center  of  the  test  sample,  load 

impedance:1M

Ω

 

Battery 

9V Alkaline battery (meet up with: NEDA

 1604A, IEC 6LR61) 

Low battery 
indicator 

>6.5V, LED is green

Overload indicator 

Red LED means the measure current is too big 
and it need to change the measuring range 

Max. insert 

10.3mm Diameter    

Dimension 

231×36×67mm 

Weight 

330g 

include battery

 

Output 

2 meters coaxial cable, terminal with insulation 

BNC connector jack

 