![multicomp pro MP720021 Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/multicomp-pro/mp720021/mp720021_manual_1854297018.webp)
User Manual
13
Caution:
The default attenuation coefficient of the probe on the instrument is preset to
10X.
Make sure that the set value of the attenuation switch in the probe is the
same as the menu selection of the probe attenuation coefficient in the
oscilloscope.
The set values of the probe switch are 1X and 10X (see
Figure 3-8 Attenuation Switch
Caution:
When
the attenuation switch is set to 1X, the probe will limit the bandwidth
of the oscilloscope in 5MHz. To use the full bandwidth of the oscilloscope,
the switch must be set to 10X.
Identify the Probe Attenuation Coefficient Automatically
The oscilloscope can identify the probe attenuation coefficient of the 100:1
(impedance 5K
±
20%) or 10:1 (impedance 10K
±
20%) probe with the identifying
pin. When you attach the probe, the oscilloscope set the attenuation automatically on
the oscilloscope vertical menu for the channel to match the probe.
For example, if you attach a 10:1 probe with the identifying pin, the screen will
prompt "The probe attenuation factor is X10", and set the attenuation to 10X
automatically on the oscilloscope vertical menu for the channel.
How to Use the Probe Safely
The safety guard ring around the probe body protects your finger against any electric
shock, shown as
Figure 3-9 Finger Guard
Содержание MP720021
Страница 1: ...i Digital Storage Oscilloscope 100MHz Model MP720025...
Страница 5: ...v Appendix B General Care and Cleaning 147 Appendix C Battery Using Guide 148...
Страница 27: ...User Manual 22 Control the vertical cursor line Switch to select the lines...
Страница 36: ...User Manual 31 Figure 4 1 Hamming window Figure 4 2 Rectangle window...
Страница 37: ...User Manual 32 Figure 4 3 Blackman window Figure 4 4 Hanning window...
Страница 38: ...User Manual 33 Figure 4 5 Kaiser window Figure 4 6 Bartlett window...
Страница 144: ...User Manual 139 7 Persist on P63 It s normal as the Oscilloscope is working hard on many more data points...