5. METHOD OF MEASUREMENT
5-1. Preparation and Caution before Measurement
●
Before measurements, be sure the meter lock function is released.
●
Set the range selector switch to “
~
mA” or “
~
A” range and set the power switch
to “BATT”.
If the pointer indicates within the “BATT” mark, the batteries are usable. If the
pointer indicates outside the mark, the batteries are exhausted. In this case,
replace the batteries with new ones.
●
Set the Zero (0) position of the pointer. If the position is incorrect, adjust it by the
pointer zero adjustment screw.
5-2. Reading Method of The Scale
No.
Function
Range
Multiplying
factor
①
Current
10mA
×
1
50mA
×
1
500mA
×
10
1A
×
0.1
5A
×
0.1
50A
×
1
②
500A
×
1
③
Voltage
~500V
×
1
④
Resistance
1K
Ω
×
1