5-2. Measurement of Voltage 
 
5-2-1. Measurement of AC Voltage 

 

                                                                                                               

Black    Red +

                                                                                             

① 

Set the power switch to "ON" position. 

② 

Connect the plug of red test lead into 
the “V” terminal and the plug of black   
test lead into the “COM” terminal. 

③ 

Set the function switch to “~V” range. 
The auto ranging mode is selected. 

④ 

Contact the tips of the test leads to the 
circuit under test. 

 

  Read the voltage value in the display.                                                           

                                                                                                                             
        Manual range mode:                                                                                  Fig. 2 
        Press the range switch to the manual range mode. 
        the instrument changes from the auto range to the manual ranging mode and   
        the range is held.    Subsequent pressing of switch causes the up-and down-ranging 
        as below. 
                               

              4V 

  40V 

→ 

400V 

→ 

600V 

→ 

400mV

↓ 

                       ↑──────────────────                                      

   

 
          To return to the auto ranging mode, press the range switch at least one second. 
 
 
 
 
  5-2-2. Measurement of DC Voltage 
 

① 

Set the power switch to "ON" position. 

  Connect the plug of red test lead into the “V” terminal and the plug of black test   

lead into the “COM” terminal. 

  Set the range selector switch to “ … V” range. The auto ranging mode is selected. 

  Contact the tips of the test leads to the circuit under test. 

 

  Read the voltage value in the display.     

 
        Manual range mode: 
        Press the range switch to select the manual range mode. The instruments changes 

from the auto ranging to the manual ranging mode and the range is held. 

        Subsequent pressing of the switch caused the up- and down-ranging as below. 
                                    400mV 

→ 

4V

 → 

40V 

→ 

400V 

→ 

600V 

↓ 

                  ↑───────────────────

                                                       

To return to the auto ranging mode, press the range switch at least one second. 
                                                                                                                                   

 

 

!

WARNING 

    Do not make measurements of power lines carrying more than 250V. 
    IN SOME CASES, POWER LINES MAY CARRY VOLTAGE SPIKES OF 

SEVERAL TIMES OF THE NORMAL SUPPLY VOLTAGE. THIS 
INSTRUMENT SHOULD NOT BE USED TO MEASURE POWER LINES. 
* The term of “POWER LINE” means the electrical circuit providing the 

power to factories, buildings and etc. 

 

Содержание M-270

Страница 1: ...t read this Instruction Manual completely and familiarize yourself thoroughly with all functions Keep this manual carefully to take out whenever you need 1 FEATURES Low cost high performance and True...

Страница 2: ...ctors or bus bars 3 SPECIFICATIONS AC conversion AC coupled true rms reading Crest factor 3 0 50 of the range 2 50 100 of the range Safety standard Meets the requirements for double insulation to IEC...

Страница 3: ...z A Manual range 40A 0 01A 2 0 rdg 8 dgt AC DC 2000A 400A 0 1A 1 5 rdg 8 dgt 2000A 1A V 50 60Hz V Auto manual range 400mV 0 1mV 1 2 rdg 5 dgt AC DC 600V rms 4V 0 001V 40V 0 01V 400V 0 1V 600V 1V Hz Fr...

Страница 4: ...owever this is not abnormal To release the zero adjustment mode press the zero adjustment switch at least 1 second 5 1 2 Measurement of AC Current Set the power switch to ON position Set the function...

Страница 5: ...to ON position Connect the plug of red test lead into the V terminal and the plug of black test lead into the COM terminal Set the range selector switch to V range The auto ranging mode is selected Co...

Страница 6: ...e COM terminal Set the function switch to range The auto ranging mode is selected Contact the tips of the teat leads to the circuit under test Read the resistance value in the display Fig 3 To select...

Страница 7: ...NORMAL SUPPLY VOLTAGE THIS INSTRUMENT SHOULD NOT BE USED TO MEASURE POWER LINES The term of POWER LINE means the electrical circuit providing the power to factories buildings and etc 5 5 Measurement...

Страница 8: ...aler If this is impossible however send the instrument directly to our sales office When mailing this instrument always pack it in its original or equivalent packing material and pack together with na...

Отзывы: