![MULTI MEASURING INSTRUMENTS HCL-1000DM Скачать руководство пользователя страница 4](http://html1.mh-extra.com/html/multi-measuring-instruments/hcl-1000dm/hcl-1000dm_instruction-manual_1853924004.webp)
* The display value may not become "0" even when there is no input current
of the conductor but it is caused from the measuring method & design of
this instrument and such phenomenon will not effect on any accuracy of the
actual measurement value.
* The instrument may get heat and or damage if the over value current is
loaded to the CT. Do not load the current more than 600A.
* Avoid the measurement at the place where 2 wires contact with CT at the
same time.
6-3. Instrument/Replacement of Batteries
1)
Confirm the power switch is OFF.
2)
Remove the battery covers by the
◯
+
driver. (Be careful not to miss the
screw cap).
3)
Install the batteries according to the indication of battery case.
4)
Set up the battery cover and make the screws tightened firmly.
5)
Put the screw cap on.
* Do not use the batteries by mixing new & old ones or used ones.
7
.
TESTING OF WITHSTANDING VOLTAGE
We recommend the testing if withstand voltage by loading 100KV for
5 minute between rod of hot stick and core of CT once a year for the
safety.
Manufactured by
MULTI MEASURING INSTRUMENTS CO.,LTD.
Akihabara Murai Bldg., 7F, 1-26 Kanda Sakuma-cho
Chiyoda-ku, Tokyo, 101-0025 Japan
TEL : 81-3-3251-7016 FAX : 81-3-3253-4278
Home Page :
http://www.mutimic.com/
E-mail :