MTT MS3716 Скачать руководство пользователя страница 4

  User's Manual 

MS3716 

Slim Plug-In First-Order Delay Signal Conditioner with Isolated Single/Dual Output 

 

Rev. 1.40 

10. Time Constant Setting 

You can use the time constant setting trimmer on the front panel to 
set a time constant for first-order delay within the range of time 
constants you have specified when ordering. 

  The trimmer has scales as shown. Use these scales as a guide 

to set the time constant. 

Percentage figures shown

(12.5% - 87.5%) are for

illustrative purposes only and

NOT printed on the product.

Note:

 

  After the setting, follow these steps to verify the time constant. 

(1)  Connect the unit as illustrated below. (Connections must be 

made with all devices powered off.) 

 
 

OSCILLOSCOPE

POWER SUPPLY

CALIBRATED
ADJUSTABLE
VOLTAGE/CURRENT
SOURCE

Example of I/O

Waveforms

 

(2)  Power on the devices and perform a warm-up for at least 30 

minutes. Note that insufficient warm-up will lead to 
inaccurate measurements. 

(3)  Apply a step input from 0% to 100% of the input range of the 

MS3716 unit

 

and observe the input and output waveforms 

on the oscilloscope. 

(4)  Read time t shown in the above example to determine the 

time constant. 

(5)  Adjust the trimmer and repeat steps (3) and (4) until you get 

a desired time constant. 

 

11. Adjustment

 

Since the unit is precisely factory-adjusted before shipment, no 
further adjustment is needed at the user’s side. Adjustment of signal 
conditioners requires technical expertise and fine craftsmanship. We 
recommend that you use adjustment service of MTT to save costs 
and ensure accuracy. Please contact MTT or its local 
representatives. Just for your reference, the adjustment procedure 
is described below.

 

11-1. Preparation

 

  For the terminal assignments, refer to section 5. 

  The wiring work should be performed with no power applied to 

the instrument. 

  Wiring should be connected to the terminal block. 

  At least 30 minutes of warm-up is required before adjustment. 

  The accuracy of measuring equipment to be used for calibration 

must be much higher than that of the instrument to be adjusted. 

  Since the ZERO and SPAN Trimmers work on both the output 1 

and output 2, adjustment should be performed on the output 1. 
(Adjustment of the output 1 has the same effect on the output 2.) 

11-2. Adjustment Procedure 

(1)  Connect the unit as illustrated below. 
(2) Adjust the voltage or current source to generate a signal 

equivalent to 0%* of the input range of the unit. 

(3)  Slowly turn the ZERO Trimmer to get 0%* output while applying 

0%* input as above. 

(4) Adjust the voltage or current source to generate a signal 

equivalent to 100% of the input range of the unit. 

(5) Slowly turn the SPAN Trimmer to get 100% output while 

applying 100% input as above. 

(6)  Repeat steps (2) through (5) until zero and span are precisely 

adjusted. 

(7)  Adjust the voltage or current source to generate 25%, 50% and 

75% of the input signal range of the unit and record the values 
of output signals to confirm they are linearly proportional. 

* For 0 to 20mA output, 0% should be interpreted as 0.5%. 
 

CALIBRATED

ADJUSTABLE

VOLTAGE/CURRENT

SOURCE

MULTIMETER

POWER SUPPLY

 

 

12. Maintenance and Inspection 

Check the unit for proper characteristics and settings about once 
every two years. 

 

13. Warranty Period and Scope

 

[Warranty Period]

 

MTT’s hardware products are warranted for a period of seven (7) 
years from the date of shipment.

 

[Warranty Scope]

 

1)  MTT warrants that its hardware products are free from defects 

in materials and workmanship and conform to its product 
specifications. 

2) MTT ships the products under its appropriate quality 

management system and quality control, but does not warrant, 
expressed or implied, that the operation, output, or indication 
of the hardware will be uninterrupted or error free. MTT shall 
not be liable and make no warranty for any damage to or any 
safety or performance trouble in the customer’s or any other 
third party’s equipment if the operation, output, or indication is 
interrupted or faulty. 

3) If, during the warranty period, the product proves to be 

defective when used in accordance with the relevant user’s 
manual, the product will be repaired or replaced. 

4)  This warranty does not apply to any malfunction or failure 

resulting from the following: 
a.  Improper or incomplete maintenance or calibration, 
b.  Any other cause, which does not relate with the delivered 

product, 

c.  Modification or repair by any person other than MTT, and 
d.  Natural disasters or other unavoidable accidents, for which 

MTT is not liable. 

 
 
 
 
 
 
 
 
 
 
 
 

 

Отзывы: