MSI-9008 CellScale Multiplexer   

   User Guide     

Page 13

M  E  A  S  U  R  E  M  E  N  T    S  Y  S  T  E  M  S    I  N  T  E  R  N  A  T  I  O  N  A  L

Setting Up and Optimizing Readings

The CellScale defaults to take 16 readings per scan of each load cell. This setting can be adjusted in order to improve
scan performance.

To adjust this setting, access the master CellScale device’s terminal
menu. From the main menu, select item one.

This will bring up the "Channel Setup" menu. From here select item
four.

This will bring up the "Channel Scan List" menu. From here select
item three.

This will bring up the "Edit a Scan List entry" menu. From here
select item six.

The CellScale will prompt for the new setting

Type in the desired setting and press Enter to accept

(1) Channels/Calibration

(4) Channel Scanlist functions

(3) Edit a scanList entry

(6) # readings per dwell: 16

Adjusting the Number of Readings Per Dwell

Enter # samples to average
before channel switch (1-255):

"Enter"

For additional information on channel setup, please reference the MSI-9000 CellScale User Guide.

Scan Order

The scan list in the CellScale is the order in which the CellScale scans the channels that are monitored. If there is a
situation in which there is a channel that requires extra monitoring, it is legitimate to add the channel into the CellScale
scan list more than once. This will ensure that as the CellScale goes through the scan list, it checks the critical inputs
more than once. Reference the CellScale owner’s manual for details.

Содержание MSI9008

Страница 1: ...MSI9008 User Guide Quality Industrial Weighing and Force Measurement Equipment Measurement Systems International CellScale Multiplexer...

Страница 2: ...D 8 Connecting Two Multiplexers to CellScale with Two A D s 9 Connecting Three Multiplexers to CellScale Two A D inputs reqd 9 Connecting Four Multiplexers to CellScale Two A D inputs reqd 9 Software...

Страница 3: ...typically a load cell is multiplexed or switched into a single channel input on the CellScale i e channel 1 x 2 x etc This signal is then measured and processed according to the setup of the CellScal...

Страница 4: ...eed through connectors are designed to accept cables with the following parameters Load Cell Cable Specifications Cable Diameter Accommodates 16 to 31 diameter cable Pressure Rating IP 68 150 PSIG 10...

Страница 5: ...SIG Negative Signal for Channel 5 7 7 Shield Shield for Channel 1 5 8 1 EXC Positive Excitation for Channels 2 6 9 2 EXC Negative Excitation for Channels 2 6 10 3 SIG Positive Signal for Channel 2 11...

Страница 6: ...The CellScale channels are summarized as follows CellScale Channel Summary Channel Function 1 ADC1 Input 1 2 ADC1 Input 2 3 ADC2 Input 1 4 ADC2 Input 2 5 Slave CellScale Channels 6 Math Channels This...

Страница 7: ...trol 0 White Gray C 4 GND Circuit Ground Orange F 5 SIG Pass Negative Signal Pass thru White Black H 6 SIG Pass Positive Signal Pass thru Green White G 7 SIG Act Negative Signal Active Output White J...

Страница 8: ...ray 3 DIG0 4 GND Orange 4 GND 7 SIG Act White Black 5 SIG Pass 8 SIG Act Green White 6 SIG Pass 9 EXC Black 9 EXC 10 Sense Brown 10 Sense 11 Sense Blue 11 Sense 12 EXC Red 12 EXC Connecting One Multip...

Страница 9: ...accomplished by connecting a Multiplexer with up to eight load cells to each CellScale input and a single load cell to the second inputs on each ADC The following diagram illustrates this This connec...

Страница 10: ...this For information about how to setup the CellScale scan list for this configuration see the Software Setup section Software Setup For details of operating the CellScale software please reference t...

Страница 11: ...ection Connecting Two Multiplexers To CellScale with Two A D s When connecting two Multiplexers to a CellScale with two ADC the primary Multiplexer will be entered in the CellScale scan list as 1 X wh...

Страница 12: ...example scan list is shown below For details about connecting the hardware for this configuration see the CellScale Connections section Connecting Four Multiplexers To CellScale Two A D inputs requir...

Страница 13: ...will prompt for the new setting Type in the desired setting and press Enter to accept 1 Channels Calibration 4 Channel Scanlist functions 3 Edit a scanList entry 6 readings per dwell 16 Adjusting the...

Страница 14: ...ems International Printed in U S A PUB 287 08 03 14240 INTERURBAN AVENUE SOUTH STE 200 SEATTLE WASHINGTON 98168 4661 U S A PHONE 206 433 0199 FAX 206 244 8470 WEB www msiscales com E mail info msiscal...

Отзывы: