MS015 COM is designed for a long-term operation 24/7. However, to ensure the test bench uptime,
it is necessary to provide regular technical inspection and below described routine maintenance
of the device.
Below are the main points that should be checked during a daily inspection:
• Environmental conditions for the tester operation (temperature, humidity, air pollution,
vibration, etc.).
• Condition of the wires that should be connected to the device outputs (visual inspection).
• Do not store or use the device in places exposed to high temperature, humidity, danger of
explosion or fire, strong magnetic field. Exposure to dampness may degrade the product's
performance.
6. TESTER MAINTENANCE
Update MS015COM software as follows:
- Download a new version of the software.
*File with the latest software version can always be found on our website: servicems.eu.
- Copy it to the root of 32 Gb MicroSD drive formatted in FAT32.
- Switch the tester off and insert the drive into the MicroSD connector of the device.
- Switch the tester on by connecting the red wire to the battery plus and the black one to its
minus, or connect the device to 12V DC power supply.
- On launch, the device will automatically detect a new version of the software and start the
installation.
•Wait until the installation is completed.
- The device will boot up upon finishing the installation.
- Switch the device off.
- Extract the MicroSD drive.
6.1 Tester software update
11
User Manual - MS015 COM Tester
English
WARNING:
Do not interrupt the software update process by disconnecting the unit or
removing the MicroSD drive.