REMOVING THE RESPIRATOR

DECONTAMINATION

DO NOT remove respirator until respirator and protec-
tive clothing are decontaminated; otherwise, exposure
to contaminants may result. Follow decontamination
and disposal procedures established by appropriate
authorities. Failure to follow this warning may result in
serious personal injury or death.

Once the protective equipment has been decontaminated,
proper disposal of affected equipment must be per-
formed. Disposal is to be performed as required by feder-
al, state, and/or local laws.

PROCEDURE FOR REMOVING THE RESPIRATOR

1. To remove the facepiece, insert the thumbs under

each of the harness head straps end tab and fully
extend the harness head straps.

2. Grasp the facepiece by the component housing or

bottom head harness straps (not the exhalation valve
or canister/cartridge(s)).

3. Pull it up and away from the face.

Note:

Before the next use, check the respirator facepiece

and if necessary, clean and disinfect. Always use a new
canister/cartridge(s). Do not reuse the canis-
ter/cartridge(s).

21

TAL 018 (L) Rev. 6 - 10073772

Содержание Advantage 4000 Series

Страница 1: ...sponsibility for using or servicing the Advantage 4000 Air Purifying Respirator The Advantage 4000 Air Purifying Respirator will per form as designed only if used and serviced according to these instructions otherwise the respirator could fail to perform as designed and persons who rely on the Advantage 4000 Air Purifying Respirator could sustain serious personal injury or death AIR PURIFYING RESP...

Страница 2: ...nfiguration as speci fied by the manufacturer O Refer to User s Instructions and or maintenance man uals for information on use and maintenance of these respirators P NIOSH does not evaluate respirators for use as surgi cal masks S Special or critical User s Instructions and or specific use limitations apply Refer to User s Instructions before donning S SPECIAL OR CRITICAL USER S INSTRUCTIONS Mers...

Страница 3: ...est Quantitative Test or Qualitative Test and follow all warnings and limitations specified 15 Wear impermeable protective clothing to prevent exposure to gases and vapors which can poison by skin absorption 16 Do not use this full facepiece with self contained breathing apparatus SCBA unless equipped with a push to connect inlet 17 Do not use this respiratory protective device in explosive atmosp...

Страница 4: ...ccordance with the instructions provided This respirator filter is NOT for use by a children or b people with a medical condition that may be adversely affected by using it Failure to follow all warnings and instructions can result in serious personal injury or death When using filters in an application that produces sparks ensure that they are protected by a shield ontact with sparks can damage f...

Страница 5: ...secup head harness and inlet assembly combined with the appropriate canister cartridge s becomes a complete respiratory protective device The respirator consists of the following subassemblies full facepiece with required nosecup assembly head harness inlet assembly filter canister cartridge s Facepiece size is identified on the front of the facepiece on the lower left head harness tab Nosecup siz...

Страница 6: ...ng poor warning properties there is no secondary means of knowing when to replace the cartridge canis ter In such cases take appropriate additional precau tions such as a more conservative change out schedule or use of an air supplied respirator or SCBA to prevent overexposure Failure to follow this warning can result in serious personal injury or death The following is a partial list of substance...

Страница 7: ...n Fluoride P100 Canister must be replaced after each use against hydrogen fluoride not to exceed 8 hours to ensure the integrity of the P100 filter F MULTI USE LIMITATIONS OptiFilter XL only 1 Approved as a high efficiency particulate air filter on approved Powered Air Purifying Respirators Also approved as a P100 filter but ONLY when used with an approved negative pressure respirator Do not excee...

Страница 8: ...NOTES 8 TAL 018 L Rev 6 10073772 ...

Страница 9: ... respirator must pass a Negative Pressure Seal Test every time the mask is donned to ensure proper fit before using the respirator The respirator must also pass the tightness test before the user attempts to enter a toxic atmosphere The respirator will not furnish protection unless all inhaled air is drawn through a suitable canister cartridge s RESPIRATOR FIT TEST The user must perform a respirat...

Страница 10: ...NOTES 10 TAL 018 L Rev 6 10073772 ...

Страница 11: ... cartridge s into the facepiece port and hand tighten If applicable place a new filter in each filter cover Never load filters into the receptacles Use the canister cartridge s immediately upon opening the bag Note For gas mask canister refer to the shelf life section for storing the canister outside the packaging Discard canister cartridge s after each use Replace the canister cartridge s before ...

Страница 12: ...NOTES 12 TAL 018 L Rev 6 10073772 ...

Страница 13: ...e harness straps 3 Insert chin into the chin cup 4 Pull the harness over the head so that the netting lies flat and is centered behind your head 5 Hold the facepiece by the inlet housing while tighten the lower straps Pull them straight back not out 6 Tighten the upper straps 7 Adjust the straps as necessary until the facepiece fits snugly against the face 8 Ensure that straps are adjusted to same...

Страница 14: ...of the strap Secure the strap in position by pulling the strap onto the button 3 Smooth the straps so that they are flat Move the slide so that it is located at the lens ring connection NEGATIVE PRESSURE SEAL TEST The Negative Pressure Seal Test must be performed each time the facepiece is donned A good face to facepiece seal must be verified before entering a hazardous area Perform the test as fo...

Страница 15: ...REPAIR Note All repair procedures assume that the respirator is clean and free of contaminant and that the fil ter s cartridge s are removed Removing the Inlet Assembly 1 Press the Inlet assem bly release button 2 Rotate the adapter clockwise a quarter turn until the notch and part number are at 9 00 o clock 3 Firmly pull the inlet assembly from the facepiece assembly by grasping the inlet assembl...

Страница 16: ... Place clip spring valve and valve stem assembly contained within rubber valve on your working sur face 9 Remove rubber valve from valve stem assembly The inlet assembly is now completely disassembled It may be cleaned using a non sudsing cleaning solution such as Confidence Plus Cleaning Solution P N 10009971 Inspect all parts for damage before reassembly Reassembling the Twin Cartridge Inlet Ass...

Страница 17: ...ocks into place 10 Roll a new O ring into position on the groove interface between the PTC adapter and the component hous ing Locate the exhalation cover 11 Place the lower tabs of the cover into the mating grooves above the adapter 12 Snap the cover into place and locate the inhalation valve disc 13 Carefully stretch the valve disc over the post in the center of the adapter The PTC Inlet Assembly...

Страница 18: ...facepiece 6 Guide the bottom lens ring over the outside of the bottom facepiece blank lens groove 7 Work the lens rings down on to the face piece blank to line up the threaded ends as close as possible with out bending the lens 8 Install one Phillips screw in the top lens ring mounting flange and thread it about halfway into the oppo site lens ring flange 9 Install the remaining Phillips screw in ...

Страница 19: ...ed 2 Installing the top straps Place the new harness on a flat surface so the straps behind the MSA logo are the top straps Ensure the black adapter clips are engaged Fold the mask rubber down against the lens Slide the slotted part of the black upper buckle up into its respective side s black adapter clip Pivot the upper buckle up and pull the strap ensur ing proper assembly 3 Install the bottom ...

Страница 20: ...NOTES 20 TAL 018 L Rev 6 10073772 ...

Страница 21: ...nt must be per formed Disposal is to be performed as required by feder al state and or local laws PROCEDURE FOR REMOVING THE RESPIRATOR 1 To remove the facepiece insert the thumbs under each of the harness head straps end tab and fully extend the harness head straps 2 Grasp the facepiece by the component housing or bottom head harness straps not the exhalation valve or canister cartridge s 3 Pull ...

Страница 22: ...NOTES 22 TAL 018 L Rev 6 10073772 ...

Страница 23: ...ns with the Confidence Plus Cleaning Solution b If the Confidence Plus Cleaning Solution is not used wash in a mild cleaning solution rinse thor oughly and submerge in a germicide solution for the manufacturer s recommended time 2 Clean and Disinfect the Facepiece a Remove the canister cartridge s from the face piece b Thoroughly wash the facepiece and nose cup in the cleaning solution A soft brus...

Страница 24: ...NOTES 24 TAL 018 L Rev 6 10073772 ...

Страница 25: ...ar for formaldehyde canister an expiration date of 6 months from the date the canister was removed from the packaging This expiration date must not to exceed the original expiration date printed on the foil packaging 4 Using the enclosed cap and plug assembly as shown in Figure A place the cap end over the threaded outlet of the canister as shown in Figure B 5 Insert the plug end on the inlet of t...

Страница 26: ...NOTES 26 TAL 018 L Rev 6 10073772 ...

Страница 27: ... the filter cover 2 Set the spacer on top of the pre filter pad 3 Fit the pre filter cover assembly over the OptiFilter HE cartridge Press firmly into place Attaching the Gas Mask Chin Canister to the Belt This conversion kit consists of Installation Instructions 1 Attach the male thread end of the breathing tube securely to the facepiece component housing 2 Slide the belt clip over the neck of th...

Страница 28: ... RING HOUSING SILICONE 9 10065803 1 NOSECUP ASSEMBLY MED LG 10065804 NOSECUP ASSEMBLY SMALL 10 10084819 1 O RING SIZE 2 024 11 10025288 2 BUTTON HEAD HARNESS ADVANTAGE 4000 SINGLE PORT FACEPIECE COMPONENTS CONT ITEM PART QTY DESCRIPTION NO NUMBER REQD 12 10065329 2 INLET ASSEMBLY 465008 1 BAG DRAWSTRING NOT 10029294 1 COVER LENS SHOWN 10075903 1 CLOTH HEAD HARNESS ASSEMBLY NOSECUP ASSEMBLY COMPONE...

Страница 29: ...ING HOUSING SILICONE 9 10065803 1 NOSECUP ASSEMBLY MED LG 10065804 NOSECUP ASSEMBLY SMALL 10 10084819 1 O RING SIZE 2 024 11 10025288 2 BUTTON HEAD HARNESS ADVANTAGE 4000 SINGLE PORT FACEPIECE COMPONENTS CONT ITEM PART QTY DESCRIPTION NO NUMBER REQD 12 10065329 2 INLET ASSEMBLY 465008 1 BAG DRAWSTRING NOT 10029294 1 COVER LENS SHOWN 10075903 1 CLOTH HEAD HARNESS ASSEMBLY NOSECUP ASSEMBLY COMPONENT...

Страница 30: ...65803 1 NOSECUP ASSEMBLY MED LG 10065804 NOSECUP ASSEMBLY SMALL 10 10084819 1 O RING SIZE 2 024 11 10018496 2 GASKET BAYONET 12 10030061 2 VALVE FACEPIECE 13 10025288 2 BUTTON HEAD HARNESS ADVANTAGE 4000 TWIN PORT FACEPIECE COMPONENTS CONT ITEM PART QTY DESCRIPTION NO NUMBER REQD 14 10065332 1 INLET ASSEMBLY 465008 1 BAG DRAWSTRING NOT 10029294 1 COVER LENS SHOWN 10075903 1 CLOTH HEAD HARNESS ASSE...

Страница 31: ...31 TAL 018 L Rev 6 10073772 ...

Страница 32: ...32 TAL 018 L Rev 6 10073772 ...

Страница 33: ...33 TAL 018 L Rev 6 10073772 ...

Страница 34: ...34 TAL 018 L Rev 6 10073772 ...

Страница 35: ......

Страница 36: ......

Отзывы: