6. Maintainance 

6.1 General 

In view of the significance of the accuracy of 
measuring results, the test device must be treated  
with care. Dirt must be avoided at all time, especially 
at the highpressure connection. 

Flushing and cleanless control of the test device 
must be made as described in para. 5.2. After 
measuring is completed, the test device shell be 
stored  in the case. The high pressure connection 
should be closed with a protective cap. 

Functual checks should be made at regular intervals 
and if erroneous measurements are suspected. 

6.2 Disorders 

If the hand wheel of the test device cannot be 
loosened while it is connected to a compressed gas 
cylinder or compressor, check if connection is under 
pressure. In this case close the cylinder valve or 
compressor valve and release air from  test device. 

Leaks at the connection to the cylinder resp. 
compressor could be due to a defect gasket. In this 
case replace the gasket. 

If the high pressure gauge or the detector tube 
holder are defective, they must be replaced. 

In case of other disorders the device should be 
returned for repair to MSA AUER or any authorized 
service station. 

 

7. Part Numbers 

MSA AUER Airtester HP 

D3188701 

(Compressed air monitoring set) 

 

Complete with case containing: 
Test device consisting of pressure reducer, high 
pressure gauge, flow meter with adjusting valve and 
water level, detector tube holder. 

Adapter for 200 and 300 bar one each for connecting 
to compressor charging panel 

Timer 

device for breaking off detector tube tips 

detector tubes for measuring carbon monoxide, 
carbondioxide, water vapour, oil - one package each. 

Accessories and Spare Parts 

Detector tubes CO-HP* 

D5085847 

Detector tubes CO

2

-HP* D5085848 

Detector tubes H

2

0-HP* D5085849 

Detector tubes Oil-HP*  

10040887 

Detector tubes Oil- HP synth.* 

D5085850 

Test device 

D4074937 

Gaskets for connector to cylinder 

D4080948 

Gaskets-lifter D4060989 

Detector tube holder 

D4074058 

High pressure gauge 

D4080073 

Timer D5185020 

Tube tip breaking device 

D5085012 

Case D5185911 

Case insert 

D5185931 

 

 * = Package of 10 pieces 

 

 

 

 

 

 

 

 

 

- 6 - 

Содержание Airtester HP

Страница 1: ...Airtester HP for compressed air according to EN 12021 Instruction for use Part Number D3188001 Edition Dec 2003...

Страница 2: ...quipment this product will do the job designed to do only if it is used and serviced in accordance with the manufacturer s instructions This manual must be carefully read by all individuals who have o...

Страница 3: ...r tube scale 3 Design The case contents all parts required for measuring see fig 1 The test device consisting of pressure reducer pressure gauge flow meter with adjusting valve and water level detecto...

Страница 4: ...w points in the flow direction of the test air therefore after inserting the tube the arrow must Break off the detector tube tips at air exit side see fig 3 Put detector tube into the tube holder inse...

Страница 5: ...point away from the test device Fig 2 Setting flow rate on flowmeter Fig 3 Breaking off detector tube tips Fig 4 Inserting detector tube into detector tube holder Fig 5 Flowmeter during Measuring 5...

Страница 6: ...sket If the high pressure gauge or the detector tube holder are defective they must be replaced In case of other disorders the device should be returned for repair to MSA AUER or any authorized servic...

Страница 7: ...7...

Страница 8: ...93 372 66 57 E Mail info msa es France MSA France 11 13 rue de la Guivernone Z I du Vert Galant F 95310 Saint Ouen L Aum ne B P 617 F 95004 Cergy Pontoise Cedex Phone 33 1 34 32 34 32 Telefax 33 1 34...

Отзывы: