Binding to a process control system: I/O modules
The IO module is a necessary module for the signal forwarding, into a control room.
This module is an interface for signal transmitting, remote operating and to read
signals, from extern transducers.
An IO module has the following features:
•
Transmit 4-20 mA output.
•
Trigger alarm outputs.
•
Reads one PT1000.
•
Reads one thermocouple.
•
Reads current inputs signals (for sensors with max. 20 mA output signals).
•
Reads voltage input signals (for sensors with max. 30 V output signals).
•
External control of the analyzer.
8.1.
Position of the IO module inside the analyzer
At the basic SWG100, one IO module is integrated. Optionally, an analyzer can be
equipped with further IO modules (max. 10).
The IO module(s) are installed on the hat-rail (position: see sketch below).
Cable gland (M16):
For cables with Ø 3,5-10 mm.
Signal cable for I/O modul
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...