MRU GmbH
SWG 100 User Manual
Page
14
of
141
2.5.
Special functions of the analyzer
H2Slow sensor protection (optional)
The instrument may be optionally equipped with a H
2
S
low
sensor. As this sensor typically shows
limited life time when experiencing high H
2
S concentrations, some precautions are taken to
protect the sensor:
•
The instrument is equipped with a cut-off solenoid valve and air purge pump to protect the
H
2
S
low
sensor, without impacting the results of all other sensors.
•
The purge unit is always activated after 3-10 minutes of active sensor measuring time.
•
The measuring time of the H
2
S
low
sensor may be limited to less than 10 minutes by user
setting
•
The H
2
S
low
sensor may optionally not be included in all measuring cycles of sampling
points. Instead the user may configure that the H
2
S
low
sensor is active only added at some of
the measurement cycles.
Содержание SWG 100 biogas
Страница 1: ...MRU GmbH SWG 100 User Manual Page 1 of 141...
Страница 2: ...MRU GmbH SWG 100 User Manual Page 2 of 141...
Страница 13: ...MRU GmbH SWG 100 User Manual Page 13 of 141 Picture above Sample gas flow diagram...
Страница 81: ...MRU GmbH SWG 100 User Manual Page 81 of 141 Device info Path EXTRAS DEVICE INFO Structure...
Страница 86: ......
Страница 92: ......
Страница 133: ...MRU GmbH SWG 100 User Manual Page 133 of 141 13 15 Option NDIR bench for 0 3 000ppm CO2 0 CH4...
Страница 134: ...MRU GmbH SWG 100 User Manual Page 134 of 141 13 16 Option Dilution unit for H2S and CO electrochemical sensors...
Страница 140: ...MRU GmbH SWG 100 User Manual Page 140 of 141 14 Declaration of conformity...
Страница 141: ...MRU GmbH SWG 100 User Manual Page 141 of 141 V01 02 EN...