MRU GmbH USER MANUAL MFplus Part 2 OPTION TRGI
31
Before starting the pipe test the pressure sensor must be selected. If the expected pressure is higher than 1,5 bar or
the medium water is used, then the external sensor must be used for the test (5 bar, 25 bar or 40 bar). The
selection of the sensors and the medium is done with the arrow buttons
;
.
With
F3
a zeroing can be done before starting the measurement.
When pressing Enter the medium (AIR or WATER) as well as the adaption time and duration time can be selected.
ATTENTION:
The internal sensor of the MFplus may never be used for tests with water -> the
device can be flooded!!
3.5.2.1
Apply pressure
F1
Aply the set pressure
3.5.2.2
Adaption phase
F3
3.5.2.3
Pressure test
F3
The actual pressure, the temperature and the
remaining time is displayed. With F3-one can
switch between the temperature- and
pressure curve.
With F1 the adaption phase can be aborted.
Once the adaption time has elapsed the
measurement will start automatically.
The measurement stops automatically once
the duration time has elapsed. With the
printer button the print out can be activated.
The measurement can be aborted with the
F2-button.