51
7.4 Details of Items Required but Not Supplied
7.4.1 Strain Gauges
The required strain gauges will need to match the delivered
Pre-Amplifier
resistance. This is
120 ohm
(standard) or
350 ohm
(requested
specially).
7.4.2 External Trigger
The user can not trigger internally from a strain channel. If only strain measurements are being made, an external trigger is required. This
can be as simple as a push button with the settings of the
DataTrap II™ Data/VOD Recorder
set to make. Refer to the external trigger
section of this manual for more details on the different trigger methods. The external trigger is not required IF the user is recording another
channel that can be used (either scope or the optional
VOD
).
7.5 Installation Of Software
Refer to the
DAS™ Data Acquisition Suite Manual.
7.6 Preparing The DataTrap II™ For A Strain Gauge Test
A
Strain Gauge
test can be set up with a pre-programmed set of values or more specific settings, including
VOD
channels, can be set up
using the normal settings software, then the values for the channels using the
Strain Gauge
can be entered later.
A
Strain Gauge
test must be set up with the following settings:
1. The
Voltage Range
must include [0 to 7.5] Volts. A wider range, such as +/- 7.5 or +/- 10 Volts can be used to allow later flexibility
of use of channels in the field if other devices might be used on some channels. If some channels are set as
VOD
, make note of
them. The setting of each channel can be checked in the field using the
LED
menu on the
DataTrap II™ Data/VOD Recorder
before
attaching cables.
2. If none of the channels are
VOD
,
Trigger Type
should be set to
External
for the channels that will have
Strain Gauges
attached.
VOD
channels can be set to internal trigger with the
Strain Gauge
channels set with no trigger – they will trigger when the
VOD
channel
triggers.
3. During the test, record which channels used a
Strain Gauge
and the
Gauge Factor Values
for these channels.
4. During the downloading of the test the gauge factors for each channel can be entered.
7.7 Field Setup Of The DataTrap II™ Strain Gauge System
The steps to collect
Strain Gauge
data with the
DataTrap II™ Data/VOD Recorder
are detailed below:
1. Turn on the
DataTrap II™ Data/VOD Recorder
and confirm that there are tests remaining on it and the battery has been charged.
2. Connect one end of the
3-pin Signal Condition Power Cable
to the DC Out port on the back of the
DataTrap II™ Data/VOD Recorder
and the other end of the 3-pin cable to the
Signal Conditioner Lid
. If the power light does not come on, check that the cable has been
connected correctly.
3. Connect the
External Trigger
cable. The power and data cables for each channel must now be set up as detailed below.
Содержание DATA TRAP II
Страница 5: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 1 Chapter 1 Getting Started...
Страница 7: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 3 Chapter 2 Introduction...
Страница 11: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 7 Chapter 3 Hardware...
Страница 25: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 21 Chapter 4 Memory And Triggering...
Страница 28: ...DataTrap II Operations Manual Edition 5 4 24...
Страница 29: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 25...
Страница 30: ...DataTrap II Operations Manual Edition 5 4 26...
Страница 31: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 27 Chapter 5 Recording VOD And Hole Deck Delay Times...
Страница 45: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 41 Chapter 6 Recording Voltage Signals From Gauges...
Страница 50: ...DataTrap II Operations Manual Edition 5 4 46...
Страница 51: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 47 Chapter 7 Strain Gauge Upgrade...
Страница 57: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 53 Chapter 8 Contacting MREL for Technical Support...
Страница 59: ...T 1 613 545 0466 F 1 613 542 8029 www mrel com 55...