and cover the plastic protective cover, put it in a dry room for avoiding the damage of parts
and affecting the use result.
11. Problem solution
Phenomenon
Reassons
Treatment
No power supply
The socket has no power
supply
Check the line and change
the socket
The socket isn
’t plug well or
the wire is broken
Change the plug and
connect the wire well
Fuse is open cirvuit
Change the fuse
The power is not switch on
Switch on the power
The temperature in the box
isn
’t rising
The setting temperature is
low
Adjust the setting temperature
The electric heater is wrong
Change heater
The temperature controller is
wrong
Change the temperature
controller
The temperature sensor
loosened
Screw the connecting nut
The error between the setting
temper ature and temperature
in the box is big
Damage of sensor
Change the temperature
controller
Temperature is out of
control
Change the temperature
controller
Fix the sensor well
The temperature controller is
wrong
Change the temperature
controller