Operating procedure 

1.

 

Press on/off (1) button to activate the pump after making sure all tubing are

well connected. 

2.

 

To setup vacuum by adjusting the regulator (5).

* Warning: When pump is used for aqueous solution filtration, please ensure the 
liquid level of flask not exceed safety level. Failure to comply could result in 
serious damage to the pump and void the warranty. If it occurs, please contact 
the distributor immediately for service. 

1.

 

If the moisture trap is full of liquid more than 70%, please stop doing the

experiments immediately and remove the liquid. 

2.

 

After finishing the experiment, please remove the liquid in the moisture trap

and pumping the air for at least 2 minutes to withdraw the residual steam. 