MRC TM-969 Скачать руководство пользователя страница 8

4)Press " Mode Button "  ( 3-5, Fig. 1) for scrolling more

display function as follows.

Here will show the emissivity data. (The
default emissivity is 0.95.)

Press Mode key ( 3-5, Fig. 1 ) for the
Maximum (MAX), Minimum (MIN), Different
between MAX and MIN (DIF) and Average
(AVG) modes. During the measurement, the
special modes reading will be displayed
beside the mode icon.
Press " Up Button " ( 3-4, Fig. 1) or " Down
Button "  ( 3-2, Fig. 1 ) key to change the
High Alarm (HAL) or Lo Alarm (LAL), then
press " Measure Button "  ( 3-6, Fig. 1 ) to
confirm it.
For example: When the reading  26.9   <

LAL 27  , the Low icon will flash and you

will hear a beep sound.
Connect the thermocouple with "
Thermocouple Socket " ( 3-11, Fig. 1 ) and
put the probe in/on the target, the
thermometer will display the temperature
automatically without pressing any button. To
see the minimum or maximum data during
the probe measurement, please hold down
the " Up Button " ( 3-4, Fig. 1 ) or " Down
Button " ( 3-2, Fig. 1).

After measure high temp, the probe
may remain HOT for a while

5) The thermometer will automatically shut off if left idle

for more than 60 seconds.

  6

Содержание TM-969

Страница 1: ...strument its durable structure developed Please read the following instructions carefully and always keep this manual within easy reach OPERATION MANUAL PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERAT...

Страница 2: ...tton 4 3 2 Down Button 4 3 3 LCD Display 4 3 4 Up Button 4 3 5 Mode Button 4 3 6 Measure Switch 4 3 7 Battery Cover 4 3 8 Infrared Lens 4 3 9 Laser 4 3 10 LED 4 3 11 Thermocouple Socket 4 4 MEASURING...

Страница 3: ...ock mode Last measurement memory Hi Lo Temperature Audio Alarm Laser target and Back light On Off control Adjustable emissivity value from 0 1 to 1 00 Preset emissivity value at 0 95 Power supply by t...

Страница 4: ...1 1 000 1 000 1 1 1 000 1 000 Circuit Exclusive microcomputer circuit Emissivity Adjustment range 0 10 to 1 00 Adjustment step 0 01 0 95 default emissivity value Infrared Temp Thermocouple pie Temp Se...

Страница 5: ...5 mA with target light on Approx DC 29 5 mA Operating 0 to 50 32 to 122 Temperature Operating Less than 80 RH Humidity Weight 386 g 0 85 LB Dimension 203 x 197 x 47 mm 8 0 x 7 7 x 1 8 inch Standard O...

Страница 6: ...PANEL DESCRIPTION 3 1 Emissivity Button 3 2 Down Button 3 3 LCD Display 3 4 Up Button 3 5 Mode Button 3 6 Measure Switch 3 7 Battery Cover 3 8 Infrared Lens 3 9 Laser 3 10 LED 3 11 Thermocouple Socket...

Страница 7: ...e Spot is 50 1 The two laser points are the reference for the target spot size Please make sure the target area is within the field of view 3 Press Emissivity Button 3 1 Fig 1 for setting the emissivi...

Страница 8: ...AL or Lo Alarm LAL then press Measure Button 3 6 Fig 1 to confirm it For example When the reading 26 9 LAL 27 the Low icon will flash and you will hear a beep sound Connect the thermocouple with Therm...

Страница 9: ...1 for laser function ON OFF With flash light be turned on in the same time as laser 5 MEASURING CONSIDERATION 5 1 LCD error message The thermometer incorporates visual diagnostic messages as follows H...

Страница 10: ...splayed when the temperature being measured is outside of the measurement range 6 BATTERY REPLACEMENT When the Low Battery icon indicates the battery is low the battery should be replaced immediately...

Страница 11: ...for many general purpose application Thermocouple Measure Range 50 to 900 Probe 50 to 1650 Type K TP 02A Dimension 10cm tube 3 2mm Dia Thermocouple Measure Range 50 to 1200 Probe 50 to 2200 Type K TP...

Отзывы: