Each digit indicate the following status :
D15
Start Word
D14
4
D13
1 ~ 9
D12 & D11
Annunciator for Display
01=
℃
02=
℉
04 = %RH
D10
Polarity
0 = Positive
1 = Negative
D9
Decimal Point(DP), position from right to the left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1
Display reading, D1 = LSD, D8 = MSD
For example :
If the display reading is 1234, then D8 to D1 is : 00001234
D0
End Word
RS232 setting
Baud rate
9600
Parity
No parity
Data bit no.
8 Data bits
Stop bit
1 Stop bit