Temperature Offsetting (OFFSET)
The temperature shown on the Actual Temperature Display is
measured by a temperature sensor inside the unit. However, this
temperature can be different from the temperature of your own
thermometer which you may use as a standard for your specific
applications. If needed, you can offset such temperature differences
at 0.1°C interval
NOTICE
If, for example, the actual temperature of the top plate is 100°C but
the displayed temperature is 98°C, you can match the displayed
temperature with the actual temperature of the unit by selecting the
offset value of +2°C
Stir Control Knob
Turn it clockwise to set desired stirring speed.
Turn it all the way counterclockwise to turn off stirring function.
Stirring Instructions
1.
Fill vessel with liquid to be stirred.
2.
Place stir bar into vessel.
3.
Place vessel in the center of the top surface.
4.
Turn Stir Control Knob
5.
Stir bar to cease rotation before removing the vessel from the
unit.
NOTICE
Please note that the viscosity of the stirred material will affect the
ability of the stir bar to remain coupled to the rotating ring magnet.
Materials of high viscosity must be stirred at slower speed
settings.