Chromatograph Series
User Manual
MRC Instrument
4
The switch of [Programmed Temp Control Setting] is to send the information of area E to the corresponding function
module. To confirm any modifications in area E, this button shall be clicked.
6.3.5 Programmed Temp Control Data Area
See area [E] in Figure 6.3.
For programmed temperature setting, when the oven is set as the programmed temperature control method, the user
can modify the contents in the area, including the holding temperature time and the next heating rate. Note that the
temperature of the next phase must be higher than the temperature of the previous one, and the user can set up to
nine-phase programmed temperature.
Note: After the data of each stage is set, the user must click the button in D area to save the data.
6.4 Detector Interface