4. Installation and testing
Users must strictly comply with the installation instructions contained in this chapter. Be advised!
Warning
⚫
Improper power connection might damage the centrifuge.
⚫
Ensure to check the power supply before connection for compliance with the requirements.
4.1 Mounting position
(1) Place this centrifuge on a firm
、
flat and level surface, ensure the four feet of this centrifuge stand on the
counter firmly. Avoid installing on a slippery surface or surface prone to vibration.
(2) Ideal ambient temperature is 20
℃
±
5
℃
, avoid placing the centrifuge in direct sunlight if temperature
(3) exceeds 30
℃
.
(4) Ensure to clear 10cm from the centrifuge on both sides and behind the centrifuge in order to ensure
air cooling efficiency.
(5) Prevent heat/water leakage near centrifuge, otherwise the rise in sample temperature or centrifuge
failure might occur easily.
4.2 Connection between power cable and ground wire
Warning
⚫
Don’t touch the power cable or power switch with wet hands to avoid electric shock.
⚫
This centrifuge must be properly grounded.
A minimum of 10A outlet providing a good grounding terminal is required and needs to meet the local
safety requirements.