background image

(2) Centrifugal chamber

If the centrifuge chamber becomes dirty, clean it using a cloth/ sponge soaked with neutral cleaning

agent.

The centrifuge chamber can be sterilized using a cloth soaked in 70% alcohol solvent.

(3) Drive shaft

It is suggested that the drive shaft should be subjected to periodical maintenance by wiping it using a

piece of soft cloth and applying a thin layer of silicone grease on it.

(4) Upper cover

Clean or sterilize the upper cover in the same manner as that of centrifuge(1).

(5) Rotor

If  the  rotor  is  put  out  of  service  for  a  prolonged  period  of  time,  please  remove  the  rotor  from  the

centrifugal chamber, remove the rotor’s lid and place the rotor upside down to dry the rotor hole and

prevent corrosion.

If any sample leaks into the rotor hole, flush the rotor hole with clean water and apply a thin layer of

silicone grease on the rotor surface after it is dried.

The  rotor  needs  periodical  maintenance.  It  is  suggested  that  the  rotor  should  be  cleaned  once  every

three months to ensure cleanliness of the tube hole and main shaft hole before applying a thin layer of

silicone grease.

9.2 Wearing parts 

The following lists the wearing parts of this centrifuge. Please replace the wearing parts in time according 

to the suggestions in the following table.   

No. 

Wearing part 

Replacement conditions 

Sealing ring of the outer lid 

Crack 

Rubber seat of temperature sensor 

9.3 Routine check 

(1) Check whether the tabletop on which the centrifuge is placed is solid, level and flat and ensure the four

feet of the centrifuge remain in contact with the tabletop. 

(2) Check whether the machine is reliably grounded, using a multi-meter. Check whether the grounding

pin in the power cable, centrifugal chamber and motor shaft are short-circuited. A short-circuit indicates a 

Caution 

Never directly pour water or other solvents into the centrifugal chamber as these solvents might enter

into the drive unit and cause corrosion or damage to the bearings.

Содержание HCEN-24

Страница 1: ...HCEN 24 Economical Capillary Centrifuge User Manual PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 2 22...

Страница 2: ...e the balanced centrifuge tube 4 7 4 Check rotor 4 7 5 Symmetrically load centrifuge tubes into rotor 4 8 Operation 4 8 1 Normal operation 4 8 2 RCF operation 7 8 3 Transient operation 8 9 Cleaning Ma...

Страница 3: ...ein are not strictly complied with NOTE Notes indicate an area or subject of special merit emphasizing either the product s capability or common errors in operation or maintenance Do not use this cent...

Страница 4: ...s placed on a solid flat and firm table Ensure that the centrifuge is on flat level table before operation Ensure that the angle between the housing and upper lid is larger than the 70 degrees while o...

Страница 5: ...the rotor is provided with a lid ensure it is secure it tightly before operation If there is any abnormality observed during operation Ensure to stop the centrifuge immediately contact the service ce...

Страница 6: ...functions Rotation speed acceleration switchover transient operating performance Current run time parameter display and sound reminder 2 Scope of application Scope of application suitable for separat...

Страница 7: ...20 5 avoid placing the centrifuge in direct sunlight if temperature 3 exceeds 30 4 Ensure to clear 10cm from the centrifuge on both sides and behind the centrifuge in order to ensure air cooling effic...

Страница 8: ...Rear view of centrifuge 6 Operating panel Fig 6 1 Schematics of operating panel Power switch Power socket Vent Sight glass Upper cover Speed display Door lock display Time display 1 3 4 2 120 x 100rpm...

Страница 9: ...ss this knob to choose the desired speed or acceleration Fig 6 2 Schematics of main display The main display is shown in Fig 6 2 in which the speed is set at 12 000rpm and time is set to 60 minutes Th...

Страница 10: ...otor on this centrifuge 7 5 Symmetrically load centrifuge tubes into rotor Caution Ensure to secure the rotor on the main shaft firmly Otherwise the rotor might fall off while the centrifuge is in ope...

Страница 11: ...plays the last operating parameters as shown in Fig 8 3 Fig 8 3 Last operation interface Speed setting 12000 rpm Time setting 60minutes The upper lid is closed 8 1 1 Rotor installation and replacement...

Страница 12: ...nob to set the selected parameters whichever parameter icon flashes Rotate the parameter adjustment knob in clockwise anticlockwise direction to increase or decrease the parameter values to achieve th...

Страница 13: ...o accumulate 3 Error message The centrifuge will automatically stop if any failure occurs in the running process with the failure code indicated on the time display window By looking up table 10 1 the...

Страница 14: ...e enters into the operation mode and displays the last set parameter values and allows to reset the target rotation speed 3 Press and hold the pulse key to increase the rotation speed until the set ro...

Страница 15: ...dried The rotor needs periodical maintenance It is suggested that the rotor should be cleaned once every three months to ensure cleanliness of the tube hole and main shaft hole before applying a thin...

Страница 16: ...cover fails The door opens during operation Press the key if the door is opened Immediately close the outer lid Close the lid before operation E 03 Wrong rotor identification Rotor can not be identifi...

Страница 17: ...k button on the right side of the machine At this time the upper cover may be turned open 2 When the centrifuge is turned OFF press the door lock button on the right side of the machine to open the up...

Страница 18: ...entrifuge tube Caution Carefully read the user manual and correctly install and use the rotor Don t exceed the maximum speed allowed by the rotor test tube and adaptor The maximum speed allowed by cer...

Страница 19: ...er drying 4 It is recommended for regular rotor maintenance and should be cleaned every 3 months to keep the tube holes and shaft clean 5 Apply a thin coat of silicon grease 11 2 Centrifuge tube 11 2...

Страница 20: ...r horizontal manner it will deform due to the effect of gravity 2 Remove the threaded cover and inner cover to prevent deformation or crack of the centrifuge tube 3 Take the centrifuge tube only when...

Страница 21: ...te of delivery Don t use any rotor damaged due to corrosion or fatigue The damage to the entire unit or rotor due to any of the following reasons is outside the scope of warranty 1 Damage due to impro...

Отзывы: