- 14 -
VI. Problems and Troubleshooting
No.
Problem
Possible Reason
Trouble Shooting
1
No power
The fuse burned
Bad connection of power
cord
Replace the fuse
Plug the power cord
again
2
Solvent
Gas
leakage
1.Insufficient condensate
2.The rubber ring at the end
of the condensate pipe is old
and can not seal the pipe
tightly
3.The condensate pipe is
broken
Put in more condensate
Replace the rubber
ring
Replace the condensate
pipe
3
System
failure
or
crash
Check if there is any strong
magnetic field or electronic
field around
Turn
off
the
instrument,move
away
the
source
of
disturbance and reboot
the system
VII. Maintenance
1. The analyzer needs heating up, so good ventilation and heat dissipation
are necessary.
2. It is recommended that in the certain period, the same extraction system
should use the same solvent to prevent cross-contamination between
different solvents.
3. Clean the instrument with a wet cloth to avoid static electricity; clean the
heating plate with an abrasive cloth to improve heat transmission.
4. Keep the receiving flasks and extraction containers clean. Clean them
)$46 User Manual