MRC DO-5510HA Скачать руководство пользователя страница 11

7)Rinsed the probe accurately with normal tap water after

each series of measurement.

5-3 Oxygen in Air ( O2 ) measurement 

1)After the meter be calibrated ( above procedure 5-1 ),

now the meter is ready for O2 measurement. 

2)Slide the " O2/DO selector " ( 4-10, Fig. 1 ) to the " O2 "

position.

3)The display will show the air oxygen in % values.

5-4 Temperature measurement

During the measurement, the lower LCD Display will 
show the temperature  values of measuring solution.
* Push the "  /  button " ( 4-4, Fig. 1 ) once to select

℃ ℉

measuring unit in   or  .

5-5 Data Hold 

During the measurement, Push the " Data Hold button "
( 4-3, Fig. 1 ) will hold the display values & LCD will
show the  " D.H " marker.
* Push the " Data Hold button " again will exit the
  data hold function.

5-6 Data Record( Max., Min., Average reading)
* The DATA RECORD function displays the

maximum, minimum and average readings. To start
the DATA RECORD function, press the " Memory
Record button " ( 4-5, Fig. 1 )  once. " REC " symbol
will appear on the LCD display.

8

Содержание DO-5510HA

Страница 1: ...you into the field of precision measurement Although this METER is a complex and delicate instrument its durable structure will allow many years of use if p r o p e r o p e r a t i n g techniques are developed Please read the following instructions carefully and always keep this manual within easy reach OPERATION MANUAL ...

Страница 2: ...the best condition when user receive the DIGITAL OXYGEN METER along the PROBE it should fill the the Probe s Electrolyte at first Probe filling Electrolyte Probe head with Diaphragm set The procedures that to fill the Probe s Electrolyte refer the chapter 8 PROBE MAINTENANCE page 11 ...

Страница 3: ...O2 Selector 4 4 11 O2 Cal button 4 4 12 MT Mountain Height button 4 4 13 Battery Compartment Cover 4 4 14 Probe Input Socket 4 4 15 Oxygen Probe handle 4 4 16 Temperature sensor 4 4 17 Probe head with diaphragm set 4 4 18 Protection cover for probe head 4 4 19 Oxygen probe plug 4 5 MEASURING PROCEDURE 5 5 1 Calibration 5 5 2 Dissolved Oxygen DO measurement 6 5 3 Oxygen in Air O2 measurement 8 5 4 ...

Страница 4: ...incorpo rated Temp sensor high precision measurement for Dissolved Oxygen DO and Temp measurement Heavy duty dissolved oxygen probe probe head can connect with BOD bottle Automatic Temp compensation from 0 to 50 for sensor Build in SALT Mountain Height compensation adjust the factor by push the button on the front panel Microprocessor circuit assures high accuracy and provides special functions an...

Страница 5: ...adjustment for best viewing angle Measurement Dissolved Oxygen 0 to 20 0 mg L liter Range Oxygen in Air 0 to 100 0 Temperature 0 to 50 Dissolved Oxygen 0 1 mg L Resolution Oxygen in Air 0 1 O2 Temperature 0 1 Accuracy Dissolved Oxygen 0 4 mg L 23 5 Oxygen in Air 0 7 O2 Temperature 0 8 1 5 Sensor The polarographic type oxygen probe with Structure an incorporated temperature sensor Probe Temperature...

Страница 6: ...ent Approx DC 6 6 mA Weight 335 g 0 74 LB batteries probe included Size Main instrument 180 x 72 x 32 mm 7 1 x 2 8 x1 3 inch Oxygen probe 190 mm x 28 mm Dia 7 5 x 1 1 Dia Accessories Oxygen probe OXPB 11 1 PC included Carrying case 1 PC Operation manual 1 PC Spare Probe head with Diaphragm OXHD 04 2 PCs Probe filling Electrolyte OXEL 03 1 PC Optional Oxygen probe OXPB 11 Accessories Spare Probe he...

Страница 7: ...ton 4 13 Battery compartment cover 4 4 button 4 14 Probe input socket 4 5 Memory record button 4 15 Oxygen Probe handle 4 6 Memory call button 4 16 Temperature sensor 4 7 Factor Adj button 4 17 Probe head with diaphragm set 4 8 Zero button 4 18 Protection cover for probe head 4 9 Salt button 4 19 Oxygen probe plug 4 10 DO O2 selector 4 ...

Страница 8: ...utton 4 8 Fig 1 and the display will show zero value 4 Connect the Oxygen probe plug 4 19 Fig 1 with the Probe input socket 4 14 Fig 1 Wait for approx 5 minutes at least until the display reading values become stable no fluctuation Push the O2 Cal button 4 11 Fig 1 and the display will show the values exactly same as 20 9 or 20 8 As the oxygen in air is 20 9 typically so use the environment air 02...

Страница 9: ...the adjusting reading reach the desired salt values Then push the Salt button 4 9 Fig 1 again to finish the adjusting procedures At this moment the new salt values will be executed c If the measured liquid is the pure water or the factor of salt can be neglected then adjust salt compensation value to 0 4 Height compensation of the probe Bear in mind that the DO measurement is considered to be take...

Страница 10: ...erature compensation to take place b As for the thermal equilibrium to occur between the probe the measurement sample must be allowed to pass which usually amounts to a few minutes if the Temp difference between the two is only several Celsius degrees 6 a In order to measure the dissolved oxygen content in any given liquid it is sufficient to immerse the tip of the probe in the solution making sur...

Страница 11: ...he lower LCD Display will show the temperature values of measuring solution Push the button 4 4 Fig 1 once to select measuring unit in or 5 5 Data Hold During the measurement Push the Data Hold button 4 3 Fig 1 will hold the display values LCD will show the D H marker Push the Data Hold button again will exit the data hold function 5 6 Data Record Max Min Average reading The DATA RECORD function d...

Страница 12: ...long with the average value will appear on the display d To exit the memory record function push the RECORD button once again The display will revert back to the current reading 5 7 Quick operation procedures Calibration procedures Power on slide the DO O2 selector to O2 position Disconnect the probe from the meter Push the Zero button display will show zero Connect the probe to the meter Until th...

Страница 13: ...etermine the Salt Height values For the general application it is not necessary to make the adjustment for the salt Height values Immerse the probe head into the measured liquid Other function DATA HOLD MEMORY RECORD RS232 output Max Min AVG Power management AUTO POWER OFF or MANUAL POWER OFF Not available in Memory Record function 10 ...

Страница 14: ...end to let the DO probe keep the best condition When user receive the DIGITAL OXYGEN METER along the PROBE it should fill the Probe s Electrolyte at first User already use the meter for a certain period Whenever user can not calibrate the meter properly or the meter s reading value is not stable please check the oxygen probe to see if the electrolyte in the probe head container is run out or the d...

Страница 15: ... Due to this characteristic the oxygen may diffuse throughout the electrolyte solution contained in the probe and its concentration may be quantified by the measurement circuit This sensitive diaphragm is rather delicate is easily damaged if it comes into contact with solid objects or is subjected to blows If the diaphragm is damaged or the electrolyte is run out it must be replaced in the followi...

Страница 16: ... the old Electrolyte from the container of the Probe head 3 Fill the new Electrolyte OXEL 03 into the container of the Probe head 4 Screw the Probe head 8 3 Fig 2 into the probe body Fig 2 8 1 Probe handle 8 2 Temp sensing metal 8 3 Probe head 13 ...

Страница 17: ...ec measurement may still be made for several hours after low battery indicator appears 2 Slide the Battery Cover 4 13 Fig 1 away from the instrument and remove the battery 3 Replace with 9V battery heavy duty type and reinstate the cover 4 Make sure the battery cover is secured after changing the battery 14 0603 DO 5510HA ...

Отзывы: