9.5.3 Data output to PC via RS-232C
In order to output data to an external computer via RS-232C, it is necessary to connect the optional
infrared RS converter (98-029-0007) as well as the data acquisition software (Mini Term). (It can be
reached from our website (
www.mrclab.com
). Click “PRODUCT,” then click
“DA-130N (Portable Model)” in the Density/Specific Gravity Meter section. The “Data communication
Software for Portable Analyzer” (98-439-0062) can be downloaded from there.) The optional stand
(98-029-0006) will help you easily set up infrared receiving unit of DA-130N.
An example of using the stand is described here in this manual. Install the main unit and infrared RS
converter onto the stand as shown below and connect the infrared RS converter to RS-232C of the
computer.
Stand
(98-029-0006)
Infrared RS converter
(98-029-0007)
Computer
Select “RS” on <Interface> of the Function screen to make communication settings.
ok
•
The optional infrared RS converter (98-029-0007) works only at 9600 baud.
Therefore, “Baud Rate” must be set at “9600.”
Output format of measurement results are shown as below:
Data
*1
Sample
No. *2
Data queue
STX
,
,
,
Byte
start=0
1
1
1 4 1
1
1
Data Date
Time
Data
queue /
/
,
:
,
Byte start=10
10
1
5
1
Data Result
Unit
Data
queue *
* * * *
*
*
,
*
*
*
*
*
*
* * * * ,
Byte start=27
7
1
10
1
Data Temperature Unit
Data
queue *
* * . *
,
,
Byte start=46
5
1
1
1
Data
α
*3
α
value
Identification
Data
queue A
, * *
*
*
*
CR CF EOT
Byte start=54
1
1
1
5
10
1
1
1
*1 : Sample name (Blank, a to z)
*2 : “*” is inserted when data is deleted. Blank except when data is deleted.
*3 : The number of
α
(0 to 9)
•
Blank from
α
to
α
value except when Comp.Density, S.G., Baume, or Conc. is selected for
results unit.