Key damaged
replace
Motor damage or leakage
replace
Running but not speed, the
machine has strange sound or
smell
Control system or motor failure manufacturers
maintenance
The machine with fault prompt, when the machine fails, the "display window" will
display the following items:
E1
Unbalance
E2
Over-speed
E3
Lid-open
E4
Lid exception
E5
Temperature sensor error
E6
Over temperature
E7
Motor stall
E8
Communication error
E9
System error
EA
Low voltage
EB
Overcurrent protection
EC
Motor fault (Hall state error)
ED
Brake resistance
Note:
The above alarm is displayed in the "display window", when appears, the buzzer
alarm for a long time short and alarm shutdown. In the event of an alarm, the machine
will be correct display only when the fault is released. If the fault is not released, press
the “Enter”to clear the buzzer alarm. At this time, the fault display will not be cleared.
When the hardware fault is solved, press the Enter key again to clear the fault display