8

press “up” and “down” button to change group number. The screen displays 
parameters settings (sample volume, sampling delay time and sampling location code) 
of the current group. Choose an appropriate group and press "Start" then the sampler 
begins to prepare sampling, meanwhile sampling delay time countdown starts, 
sampling pump will run (begin sampling) when the countdown finished. After 
reaching to the set sampling volume, the pump will automatically stop, and the screen 
will return to the main interface, sampling finish, sampling data will be automatically 
saved. 
(5) Remove the sampling head, take out the dish and cover it with its original cover, 
label or number it if necessary, then place it in a thermostatic incubator which has 
reached the set temperature for incubation. 
(6) If change location for another sampling process, it need to sterilize again, repeat 
the previous steps. 
(7) In the main interface, press "Select" to data query, you can see the previously 
recorded sample data (including sampling quantity, sampling date, sampling time, 
sampling location code and other information), it can save 2000 pages of data, press 
"Select" to adjust the cursor position, press “up” and “down” button to turn page, 
press "Exit" to return to the main interface. 
(8) During sampling, if press "Exit" will manually terminate sampling, the pump will 
stop working and the display of the total sampling volume will stop, the screen returns 
to the main interface. 

VIII. Maintenance and precautions

1. Shall not sampling toxic or corrosive gas, not use it in high acidity and alkalinity
environment, not use it in dust concentration more than 0.3mg/ m 

3

 environment. 

2. Do not start sampling before removing the cover in case of pump get overheat or
burnt. 
3. If display shows low power (empty in the batter symbol) it indicates the battery
must to be charged, otherwise it will affect the battery life of the sampler. 
4. When installing dish, dish must be in place and fastened, and adjust the dish to the
approximate water level. 
5. After each use, must to use alcohol to swab clean "sampling head" and "protective
cover", and put the "protective cover" on the "sampling head". 
6. If not use for a long time, it should be placed in the box. We recommend use it at
least once per month, each time running 20 minutes. 

Содержание AIS Series

Страница 1: ...AIS SERIES Microbial Air Sampler Instruction 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 5 18...

Страница 2: ...equipment please keep it for future reference Directory i Equipment introduction 2 ii Application areas 2 iii Working environment 2 iv The main features 3 v Technical Parameters 3 vi Structural descr...

Страница 3: ...ium will maintain the activity of the microbial by creating the best environment which ensures accurate and reliable measurement results After automatic sampling is completed place the Petri dish in t...

Страница 4: ...ctricity 7 Built in high capacity rechargeable lithium polymer battery its fully charged continuous working time is more than 15 hours 8 Comply with ISO DIS 14698 1 sampling criteria 9 Can retrieve re...

Страница 5: ......

Страница 6: ...g location see detail in the function menu panel Save sampling setting Function Select a saved location number Inquire past sampling record Start Start Menu Select Delete sampling setting Delete sampl...

Страница 7: ...ress the up and down button to adjust the sample volume in the range of 10 to 6000 liters adjusted in steps of 10 liters press Exit to exit parameter setting press Menu to enter the next parameter set...

Страница 8: ...installation and testing According to GMP requirements to choose culture medium for targeted planktonic microbial and prepare the required ready to use plate According to the requirements of GMP choo...

Страница 9: ...ng time sampling location code and other information it can save 2000 pages of data press Select to adjust the cursor position press up and down button to turn page press Exit to return to the main in...

Страница 10: ...place a new fuse The pump not work when press Start for sampling 1 Wire is broken inside the handle 2 Pump power fail 3 Pump fail 1 Reconnect the wires inside the handle 2 Replacement pump power 3 Re...

Страница 11: ...rance 2 Function tests 3 Battery performance check 4 Parametric tests Sampling Location Map a b c d e Average Sampling head wind 0 38m s Sampling time error 0 5S Measuring tube size 75 0 1 5 Conclusio...

Отзывы: