Bump Test 

Bump test is to check if the sensors and alarms are 

still working properly. 

 

To enter bump test: 

        Config mode 

 Sensor Menu 

 Bump Test. 

 

To bump test the unit: 

        Select Sensor 

 Apply Gas 

 Start Bump 

 

Wait Count-down Finish 

 Get Result. 

 

 

 

 

 

 

 

 

 

Fresh Air Calibration 

Fresh air calibration is to get the baseline for sensor, it 

is done in normal fresh air.   

 

To enter fresh air calibration:   

Config Mode 

 Sensor Menu 

 Fresh Air Calib. 

 

To change sensor selection:   

All sensors are selected by default, press Right 

Key to start selecting, toggle selection with Left 

Key, move high-light with Right Key, press “OK” 

when selection is done. 

 

Start fresh air calibration: 

        Press Left Key to start calibration, a count-down 

timer starts. User can cancel calibration during 

count down. Calibration result will be displayed 

on LCD screen after count-down is finished, 

“Pass” or “Fail”. 

Span Calibration 

Span calibration is to get the metric of the sensor to 

the gas, it is done with a known concentration gas. 

 

To enter span calibration: 

        Config Mode 

 Sensor Menu 

 Span Calib. 

 

To calibrate the unit: 

        Select Sensor 

 Apply Gas 

 Start Calibration 

 Wait Count-down Finish 

 Get Result. 

 

To apply gas: 

Apply gas to gas inlet, on the top of the unit.

 

 

 

 

 

 

 

 

10 

Maintenance and Service 

Battery, sensor replacement and other maintenances or 

services, please contact mPower’s service department 

or refer to the User Manual. 

Year of manufacturing 

To identify the year of manufacturing, please refer to 

the serial number of the instrument. 

The fifth to sixth digit in the serial number indicates 

the year of manufacturing. 00~99 

indicates the manufacturing year is 2000 to 2099. 

 