Care and maintenance
32
Replacement of the reverse osmosis module
7.5
-
Turn off the system by unplugging the power adapter out of the electrical outlet
-
Disconnect the system from feed water
-
Open the system
-
Remove reverse osmosis housing from its holder
-
Remove tubes from the reverse osmosis housing (caution: water leakage)
-
Open the housing and remove the reverse osmosis module
-
Unpack the new reverse osmosis module and put it into the housing
-
Close the housing
-
Reconnect the tubes onto the reverse osmosis housing
-
Plug the reverse osmosis housing back into its holders
-
Connect the system to feed water
-
Turn on the system by plugging the power adapter into an electrical outlet
-
Flush the reverse osmosis module by start the tool “CHANGE ProPak.”
(see chapter 5.2.1)
-
Check impermeability on all connections
-
The system is ready for operation