Normal Test
17
5 Normal Test
5.1 Without Moisture Sensor
1. Input settings as described on page 15 for the following:-
−
Test Reference
−
Test Type - select Normal
−
Test Number
−
Format
2. Press OK button to return to Remote Device dialogue.
3. Press Flush button to open flush valve - flush indicator turns yellow colour to indi-
cate valve is open. Leave valve open until at least 200ml has passed to waste. Refer
to page 36 for more information about flushing times.
4. Press Stop button to close flush valve.
5. Press Start button. The PML² will now commence the sampling cycle.
6. The completion bar indicates progress of the test and the status is shown as Sam-
pling.
Figure 1
A Test in Progress