26
PC Software Operation
11 PC Software Operation
The
Remote Control
dialogue allows an operator to manually control the ICM from
a PC, using the LPA-View software. It can also be used to download test results
that have accumulated during autonomous (disconnected) operation.
Figure 11.1
The Remote Control dialogue
To perform a test, first optionally edit the
Test Reference
and press
Apply
to set the
new value. This is a descriptive label which can be used to identify or group the
test later (along with the test number and test time/date). An example would be a
machine number or customer name. The
Test Reference
can be up to 15 characters
in length.
When connected the ICM
status
should show "Ready’’. The operator can then press
the
Start
button to begin the test. The
progress bar
shows how much of the test has
been completed. The test can be abandoned at any time by pressing the
Stop
button.
If the
Start
button is pressed during a test, then the current test is abandoned and a
new one started.
When the test has finished, the
Result
area will display the contamination level in
the set format and water content and temperature if ’W’ option.
Some items may be missing depending on the options fitted to the ICM.
Содержание ICM
Страница 1: ...ICM Inline Contamination Monitor 2 0 User Guide 201 008 EN www mpfiltri co uk ...
Страница 6: ......
Страница 18: ...18 Front Panel Operation Figure 5 7 History Screen General Modbus CAN Bus Figure 5 8 Diagnostic Screens ...
Страница 47: ...Electrical Interface 47 Figure 14 6 ...
Страница 55: ...Fault Finding 55 DO NOT USE ACETONE ...
Страница 82: ......