Fault Finding
Value
Function
Comment
0
NOT READY
Unit is powering-up, or there is some
problem
1
READY
Ready to start a test
2
TESTING
Test in progress
3
WAITING
Waiting between tests
128
FAULT OPTICAL
LED failure / sensor blocked / filled
with air
129
FAULT FLOW LOW
Flow too low for reliable test
130
FAULT FLOW HIGH
Flow too high for reliable test
131
FAULT LOGGING
Fault with data logging
132
FAULT WATER
SENSOR
Water sensor failure
Table 16.1
The
TEST STATUS
Register
16.3 Other Faults
Unexpected results
obtained from sample
Check that the Mini-mess hose has been fully connected
at both the system and ICM ends.
Confirm that the flow through the ICM is within the
range of the unit.
High water / aeration levels.
Remote Device dialogue
not responding to
buttons being pressed.
Check that correct COM port has been selected in the
Remote Device dialogue.
Check USB driver has been installed.
Disconnect power supply to ICM and then reconnect it.
If the ICM has been subjected to excessive contamination and a blockage is sus-
pected, a flush with a suitable solvent may clear the blockage.
The standard ICM is fitted with Viton seals, so Petroleum Ether or Iso Propyl Al-
cohol may be used for this purpose, in conjunction with the MP Filtri UK Bottle
Sampling Unit. See flushing guidelines on supplied product CD.
Содержание ICM
Страница 1: ...ICM Inline Contamination Monitor 2 0 User Guide 201 008 EN www mpfiltri co uk ...
Страница 6: ......
Страница 18: ...18 Front Panel Operation Figure 5 7 History Screen General Modbus CAN Bus Figure 5 8 Diagnostic Screens ...
Страница 47: ...Electrical Interface 47 Figure 14 6 ...
Страница 55: ...Fault Finding 55 DO NOT USE ACETONE ...
Страница 82: ......