Measure Range: 0.00~14.00PH 

 

Resolution: 0.01PH  

 

Accuracy: ±0.1PH  

 

Operating Temperature: 0°C ~ 50°C 

 

Automatic Temperature Compensation: 0°C ~ 50°C 

 

Power Supply: 4 x 1.Sv (AG-13 Button Battery) 

 

Calibration: Manual. 

 

Dimensions: 188mm x 35mm x 35mm 

 

Weight: 98g 

 
 

Operating Instructions:

 

 
 

Remove the protective cap. 

 
 

First rinse the electrode with distilled water, and suck it with filter paper.  

 
 

Turn the meter on by pressing the ON key.  

 
 

Immerse the pH meter elec-trode in the solution to be tested.  

 
 

Stir gently and wait for the reading to stabilize.  

 
 

Press the "HOLD" key, the reading is be kept. When re-peat measuring, 

 

please press the "HOLD" key again, it automatically relieves the hold 

 

condition. 

  
 

After use, rinse the electrode with water to minimize con-tamination. Turn the 

 

meter on by pressing the OFF key. 

 
 

Always replace the protective cap after use. 

 

PH Calibration:

 

 
 

Immerse the electrode in pH 6.86(or pH7.00) standard buffer solution of 

 

phosphate, and gently shake it. 

 
 

Allow the reading to stabilize and with a small screwdriver turn, the calibration 

 

trimmer beside the battery until the dis-play shows is corresponding the 

 

calibration solution. Rinse the electrode with distilled water, and suck it with 

 

filter paper. 

 