10
•
Close the condenser diagram down completely
– moving lever to C.
•
Fully open the
field of diaphragm and
remove the
diffuser filter slide.
•
Loose
n, don’t remove the lamp socket clamp screw using the knob. This will allow the knob to move
around with little resistance.
•
Holding the lamp socket clamp screw, move the lamp filament image to the center of the condenser
diaphragm iris. The filament will be visible looking down from above through the top of the condenser.
Once centered tighten down the lamp socket clamp screw.
•
After finishing the above lamp centering procedure, re-insert the filter slider.
Lamp Socket Clamp
Screw Knob