background image

11 

 

 

1.   Slightly close the diaphragm to a smaller aperture                                                            

 

2.   Objective lens is dirty. See “Care and Cleaning” Section. 

 

3.   Eyepiece is dirty. See “Care and Cleaning” Section. 

D.  PROBLEM 

– Dust or hairs seem to be moving in the image. 

 

CORRECTION – 

The iris diaphragm is not open wide enough. Slowly 

 

open the diaphragm to increase the size of the opening allowing 

 

for additional illumination. 

 

E.    PROBLEM 

- Once the specimen is in focus, it moves out of focus. 

       

CORRECTION 

– Gravity is causing the stage to drift downward,      

       causing loss of focus. The focusing tension should be increased to   
       act as a brake to prevent this downward drift 
 

F.  PROBLEM

 – Focusing knobs turn with difficulty. 

 

CORRECTION

 –  

       1. Loosen the tension collar adjustment.  See page 10. 
       2. The microscope should be disassembled, cleaned and   
           re-lubricated by a qualified, authorized technician. 

 
 
 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

Содержание M3600 Series

Страница 1: ...M3600 Series Instruction Manual WWW MOTIC COM...

Страница 2: ...ries Stage Clips Camera Port Head Arm Cord Holder Coarse Focus Knob Fine Focus Knob Mechanical Stage Control Light Intensity Control Eyepiece Nosepiece Objective Specimen Holder Stage Iris Diaphragm I...

Страница 3: ...vide full illumination to the specimen plane and to enhance the resolution and contrast of the object being viewed The condenser can be raised and lowered for precise light control CORD HOLDER Corded...

Страница 4: ...permit maximum resolution All objectives are color coded and parfocaled The 40XRD and the 100XRD are sealed to keep oil residue from seeping into the objective These two objectives have a spring load...

Страница 5: ...of an objective s light gathering capabilities The concept may be compared to the F valve in photographic lenses Generally speaking objectives with N A values of less than 1 00 are Dry objectives Valu...

Страница 6: ...g knob slowly and carefully refine the focus to clearly observe the fine details of the specimen 6 If the image of the specimen appears pale the aperture of the iris diaphragm should be slowly closed...

Страница 7: ...all and is critical to focusing the specimen Good quality glass not plastic cover slips should be used If their thickness is over 0 17mm the objective will not resolve properly because the specimen ca...

Страница 8: ...MASTER scale in millimeters For example the length is 1 0mm from the top of the arrow to the bottom the total length of the scale is 8 0mm etc The thickness of all the horizontal lines is 0 01mm 2 To...

Страница 9: ...certain components should be cleaned frequently to ensure ease of viewing The power switch should be turned off or the microscope should be unplugged when not in use CLEANING The front lens of the ob...

Страница 10: ...enter of the stage to prevent the moveable arms of the mechanical stage from being damaged during storage in the cabinet LED REPLACEMENT To prolong the life of the bulb you should always turn off the...

Страница 11: ...se the size of the opening allowing for additional illumination E PROBLEM Once the specimen is in focus it moves out of focus CORRECTION Gravity is causing the stage to drift downward causing loss of...