MOS MicroSet MS pH 97 Скачать руководство пользователя страница 11

11 

 

 

8. SLOPE 

 

 

 
Slope  indicates  health  of  pH  Sensor.  Generally  slope between 

70  and  115%

 is 

acceptable. If slope is below 70% or above 115%, you need to change the pH Sensor. 
 

 

 

9. SETPOINT 

1.  You can use SPL1 & SPH1 for 

Low range controlling of pH

 

If you enter SPL1 = 3.00 & SPH1 = 4.00, 
Then in this case Relay will get ON when pH value goes below 3.00 pH & Relay will get OFF 
when pH value goes above 4.00 pH 
 

2.  You can use SPL2 & SPH2 for 

High range controlling of pH

 

If you enter SPL2 = 7.00 & SPH1 = 8.00, 
Then in this case Relay will get ON when pH value goes above 8.00 pH & Relay will get OFF 
when pH value goes below 7.00 pH 
 
 

 

Denotations : 

 
SPL1  

=      SetPoint Low for RL-1 

SPH1       

=      SetPoint High for RL-1 

SPL2        

=      SetPoint Low for RL-2 

SPH2       

=      SetPoint High for RL-2 

 

Содержание MicroSet MS pH 97

Страница 1: ...MANUAL MS pH 97 pH INDICATING CONTROLLER TRANSMITTER...

Страница 2: ...2...

Страница 3: ...3 TECHNICAL SPECIFICATION 06 4 TERMINAL CONNECTIONS 07 5 KEY DETAILS 07 6 CALIBRATION PROCEDURE 08 7 TEMPERATURE COMPENSATION 10 8 SLOPE 11 9 SETPOINT 11 10 MODBUS RS485 optional 14 11 SAFETY WARNING...

Страница 4: ...the back plate of the instrument Warranty Terms The terms of this instrument that are warranted are described in the techno commercial offer We will make any repairs that may become necessary during...

Страница 5: ...er may only be installed commissioned operated and maintained by properly trained and authorized personnel Look at the ordering detail to ensure that the device is delivered according to your order Ch...

Страница 6: ...2 RTD PT 100 Temp Sensor through Terminals 3 4 Supply 230V AC 50 Hz Output 4 20 mA DC MODBUS RS485 Optional Temperature Compensation Manual Auto through PT 100 Range 0 149 9 O C Accuracy 0 01 pH Cali...

Страница 7: ...Sensor to BNC Connector on back plate 3 Connect RTD PT 100 Temperature Sensor if required to Terminals 3 4 on back plate Please connect RTD PT100 Temperature Sensor AFTER Calibration Procedure 4 Conne...

Страница 8: ...starting Calibration Procedure make sure Temperature is in Manual Mode at 25 C When RTD PT100 Temperature Sensor is NOT connected Temperature is in Manual Mode Remove Protective cum storage cap of pH...

Страница 9: ...9...

Страница 10: ...troller Transmitter MS pH 97 automatically goes in Automatic Temperature Compensation Mode When RTD PT 100 Temperature Sensor is Not Connected pH Indicating Controller Transmitter MS pH 97 is in Manua...

Страница 11: ...4 00 Then in this case Relay will get ON when pH value goes below 3 00 pH Relay will get OFF when pH value goes above 4 00 pH 2 You can use SPL2 SPH2 for High range controlling of pH If you enter SPL2...

Страница 12: ...12 Similarly for setting of Setpoint SPH2 SPL2 SPH2 please refer below flow chart...

Страница 13: ...13...

Страница 14: ...Indicating Controller Transmitter MODBUS communication is based on RS485 RTU User can set the salve ID as per the requirement with following Specification Baud Rate 9600 Slave ID 1 default Start Bit...

Страница 15: ...external hazards customer should take necessary care while preparing site ready before installation 4 Ensure proper supply voltage 230 VAC with proper polarity to the instruments before Powering ON in...

Страница 16: ...16...

Страница 17: ...roSet Model MS pH 97 Serial No __ __ __ __ __ Input pH Simulator pH Sensor Model MS pH ___ SN __ __ __ __ READING Standard Buffer Solution pH Observed Reading Before Calibration pH Observed Reading Af...

Страница 18: ...18...

Страница 19: ...e company reserves the right to decide the workplace for the repair work The freight for defective material will have to be borne by the buyer the transit risk for such material will rest with the buy...

Отзывы: