NAMES OF COMPONENTS 

 

LCD Display   

  Panel 

  H Button: This Button is used 

to the switch of data hold. 

  Transform Switch:   

This switch is used to select 

functions and desired ranges. 

  10A Jack: Input terminal for   

current 0 ~ 10A. 

  COM Jack: Common terminal 

for measurement. 

  hFE Jack: This Jack is used   

for transistor test. 

 

V/Ω/mA Jack

: Input Jack 

 

SPECIFICATIONS 

Accuracy is specified for a period   

of 1 year after calibration and at   

18

  to 28

℃(

64

°

F to 82

°

F

 

with relative humidity to 75%. 

 

 

GENERAL SPECIFICATIONS 

-  Max. Voltage Between Terminals And Earth Ground: 1000V DC or 

750V AC 

- Fuse Protection: F 200mA/250V (quick acting). 

- Display: 15mm LCD 

- Max. Show Value: 1999 (3-1/2) 

- Polarity Indication: ‘-’indicates negative polarity. 

- Overrange Indication: Display ‘1’   

- Sampling Time: approx. 0.4 second 

- Low Battery Indication: ‘

’ displayed 

- 4 - 

- Power Supply: 1.5V×3 AAA battery. 

- Operating Temperature: 0

  to 40

℃(

32

°

F to 104

°

F

 

- Storage Temperature: -10

  to 50

℃(

10

°

F to 122

°

F

 

- Dimension & Weight: 158×74×32 mm, 250g

including battery

 

 

 

ELECTRICAL SPECIFICATIONS 

Circumstance Temperature: 23

 

±

5

, Relative Humidity: < 75% 

DC Voltage 

Range 

Resolution 

Accuracy 

200mV 

0.1mV 

±

0.5% of rdg + 2 digits

 

2V 

0.001V 

±

0.5% of rdg + 3 digits

 

20V 

0.01V 

±

0.8% of rdg + 3 digits

 

200V 

0.1V 

 

1000V 

1V 

±

0.8% of rdg + 5 digits

 

Input Impedance: 1MΩ

 

-  Overload Protection: 200mV range: 250V DC or AC rms, 2V-1000V 

ranges: 1000V DC or 750VAC rms. 

- Max. Input Voltage: 1000V DC 

AC Voltage 

Range 

Resolution 

Accuracy 

200V 

0.1V 

±

1.2% of rdg + 10 digits

 

750V 

1V 

 

Input Impedance: 1MΩ

 

- Overload Protection: 750V AC rms or 1000V DC   

- Frequency Range: 40 to 400Hz   

- Response: Average, calibrated in rms of sine wave. 

- Max. Input Voltage: 750V rms AC 

Diode 

Range 

Resolution 

Function 

 

0.001V 

Display: read approximate forward 

voltage of diode 

-  5  -    

 