NAMES OF COMPONENTS
⑴
LCD Display
⑵
Panel
⑶
H Button: This Button is used
to the switch of data hold.
⑷
Transform Switch:
This switch is used to select
functions and desired ranges.
⑸
10A Jack: Input terminal for
current 0 ~ 10A.
⑹
COM Jack: Common terminal
for measurement.
⑺
hFE Jack: This Jack is used
for transistor test.
⑻
V/Ω/mA Jack
: Input Jack
SPECIFICATIONS
Accuracy is specified for a period
of 1 year after calibration and at
18
℃
to 28
℃(
64
°
F to 82
°
F
)
with relative humidity to 75%.
●
GENERAL SPECIFICATIONS
- Max. Voltage Between Terminals And Earth Ground: 1000V DC or
750V AC
- Fuse Protection: F 200mA/250V (quick acting).
- Display: 15mm LCD
- Max. Show Value: 1999 (3-1/2)
- Polarity Indication: ‘-’indicates negative polarity.
- Overrange Indication: Display ‘1’
- Sampling Time: approx. 0.4 second
- Low Battery Indication: ‘
’ displayed
- 4 -
- Power Supply: 1.5V×3 AAA battery.
- Operating Temperature: 0
℃
to 40
℃(
32
°
F to 104
°
F
)
- Storage Temperature: -10
℃
to 50
℃(
10
°
F to 122
°
F
)
- Dimension & Weight: 158×74×32 mm, 250g
(
including battery
)
●
ELECTRICAL SPECIFICATIONS
Circumstance Temperature: 23
℃
±
5
℃
, Relative Humidity: < 75%
DC Voltage
Range
Resolution
Accuracy
200mV
0.1mV
±
(
0.5% of rdg + 2 digits
)
2V
0.001V
±
(
0.5% of rdg + 3 digits
)
20V
0.01V
±
(
0.8% of rdg + 3 digits
)
200V
0.1V
1000V
1V
±
(
0.8% of rdg + 5 digits
)
-
Input Impedance: 1MΩ
- Overload Protection: 200mV range: 250V DC or AC rms, 2V-1000V
ranges: 1000V DC or 750VAC rms.
- Max. Input Voltage: 1000V DC
AC Voltage
Range
Resolution
Accuracy
200V
0.1V
±
(
1.2% of rdg + 10 digits
)
750V
1V
-
Input Impedance: 1MΩ
- Overload Protection: 750V AC rms or 1000V DC
- Frequency Range: 40 to 400Hz
- Response: Average, calibrated in rms of sine wave.
- Max. Input Voltage: 750V rms AC
Diode
Range
Resolution
Function
0.001V
Display: read approximate forward
voltage of diode
- 5 -