72
Regular Inspection
•
Maintenance and inspection are generally consider to be the duty and obligation of the user, but if, for
some reason, the user is unable to carry out these duties, they may be performed by the accedited
service personnel. Contact your local dealer or J. MORITA OFFICE for details.
•
This unit should be inspected for all the items in the following list once a year.
•
At the start and end of each business day, make sure that switching the main switch to ON and OFF
turns the equipment on or off without fail.
•
The inspection items marked * may only be performed by the service personnel for furether preventive
inspection and maintenance during the life of the device.
•
For repair or other types of service contanct your local dealer or J. MORITA OFFICE.
Regular Inspection List
Power Supply and Physical Stability
1. Power Supply Voltage
* Use a digital or analog tester to measure the unit’s power supply.
The result must be 120 V AC ± 10 % for EX-1, and 220/230/240 V AC ± 10 % for EX-2.
2. Ground connection
Visually inspection the ground connection to make sure it is securely and properly connected.
3. Floor and base securing bolts
Visually inspect the floor and base securing bolts.
Check that the floor is level and make sure the base bolts have not loosened up.
4. Bolt and screw tightness
Inspect all the bolts and screws on the unit.
Make sure that all bolts are in place and properly secured.
5. Electrical circuitry
Make sure all wiring and connections are intact. * Inspection of the wiring inside the enclosures may only
be done by the accredited service personnel.
6. LAN cables
Make sure no cables are bent or pinched and that they are all securely connected
7. Outer appearance and labels
Make sure outer covers and panels etc are not damaged, cracked, loose or dirty.
Make sure labels for rating, tube, and lasers are all properly in place and securely attached.
8. Main switch
Turn the main switch on and off and make sure the main LED on the control box lights up.
9. Emergency Stop Switch
Turn on the main switch and then press the emergency stop switch.
Check that the power goes off. (Main LED on the control box goes out.)
10. Patient Auto Positioning (AF)
Put a test object in place. Press the patient type keys and make sure the arm moves forward or backward.
Repeat this test 3 times with the test object in a different position each time.
Make sure auto positioning sensor moves up and down smoothly.
11. Light Beams
Make sure the Mid-sagittal, Frankfort plane and Image layer beams light up and can be turned on and off.
Also make sure that the Mid-sagittal, Frankfort plane and Image layer beams move smoothly.
12. Patient Handles
Make sure handles are tight and properly secured.