4 Cephalo Exposure
1
2
Instructions for Use 2019-03-22
Open the ear rods plates carefully with both hands to retract the
ear rods from the patient’s ear orifice.
Guide the patient away from the unit.
Remove the ear rods.
Disinfect the ear rods by wiping it with ethanol (70 vol% to 80
vol%).
• Move the ear rod plates very carefully to get the ear rods out of the ear orifice
and make sure the ear rods are all the way out. Otherwise, the patient could be
injured.
4.3.4 Patient Egress
●
PA and 45° Angle Exposures
Press the In/Out switch to
move the cephalo slit to the
right edge.
●
Lateral Exposure
Pull the nasion plate out and
then swing it up.
2. Remove ear rods
1. Guide patient away from the unit