Page 4
Operation 12. 21. ’04
PROHIBITION
•
Do not use this unit in conjunction with an electric scalpel or on patients who have a
pacemaker.
•
Do not use this unit in the medical operation room.
•
Blocked canals cannot be accurately measured.
•
This unit must not be connected to or used in combination with any other apparatus or
system. It must not be used as an integral component of any other apparatus or system.
J. Morita Mfg. Corp. will not be responsible for accidents, equipment damage, bodily
injury or any other trouble which results from ignoring this prohibition.
•
Illumination devices such as fluorescent lights and the Film viewer which use an inverter
can cause ROOT ZX II to operate erratically. Do not use ROOT ZX II near devices such
as these.
•
Electromagnetic wave interference could cause this unit to operate in an abnormal,
random and possibly dangerous manner. Cellular phone, transceivers, remote controls
and all other devices which transmit electromagnetic waves located inside the building
should be turned off.