Dental root canal measurement and treatment unit 

 
 
 
 

Low Speed Handpiece Module 

   
 

Operation Instructions

 

 

 

 
 
 
*  Low Speed Handpiece Module must be connected to Canal Measurement Module, 

which is sold separately.    This unit cannot be used as an independent unit. 

 
 
 
 
 
 
 

 

 

 

USP. 5980248 

USP. 5897315 

USP. 5902105 

USP. 5096419 

USP. 5211556 

USP. 5295833 

DBP. PAT. 4126753 

DBP. PAT. 4139424 

DBP. PAT. 4232487 

KOREA PAT. 159070 

JAPAN PAT. 2873722 

JAPAN PAT. 2873725 

JAPAN PAT. 3071245 

JAPAN PAT. 3071270 

JAPAN PAT. 3113109 

JAPAN PAT. 3113095 

JAPAN PAT. 3213484 

 

This manual is for Low Speed Handpiece Module.     
To measure a root canal refer to the manual for Canal Measurement Module. 

Содержание ROOT ZX II

Страница 1: ...be used as an independent unit USP 5980248 USP 5897315 USP 5902105 USP 5096419 USP 5211556 USP 5295833 DBP PAT 4126753 DBP PAT 4139424 DBP PAT 4232487 KOREA PAT 159070 JAPAN PAT 2873722 JAPAN PAT 287...

Страница 2: ...uick and easy reference This manual contains essential safety information CONTENTS Page 1 Prevent Accidents 1 2 Parts Identification 5 3 Assembling the Unit 6 4 Before Using the Unit 8 5 Operating the...

Страница 3: ...ry of the patient or operator if the instructions are not followed properly PROHIBITION The user can not use in such a way that may result in serious injury of the patient or operator NOTE This alerts...

Страница 4: ...rm it is securely held in place If the file is not securely placed it could come out and injure the patient see page 9 Take care to avoid the risk of a file being swallowed by the patient see page 9 C...

Страница 5: ...rmities or damage before using them Any type of deformity could result in the file breaking see page 19 If the file touches the oral mucosa or a tooth it will automatically start to rotate and could i...

Страница 6: ...Morita Mfg Corp will not be responsible for accidents equipment damage bodily injury or any other trouble which results from ignoring this prohibition Illumination devices such as fluorescent lights a...

Страница 7: ...ment unit is called Low Speed Handpiece Module Accessories Micromotor 1 Contra Angle 1 Handpiece Cord 1 TR800 Blue Line TR400 Green Line Option Handpiece Rest 1 AR Oil 1 AC Adapter 1 Battery 1 Foot Sw...

Страница 8: ...n Low Speed Handpiece Module with the notches in Canal Measurement Module and put them together 4 Slide Low Speed Handpiece Module all the way down until it is securely attached NOTE If the catch on t...

Страница 9: ...pter is connected 2 The Amber Charge LED on the back of Low Speed Handpiece Module starts flashing on and off and then after a few seconds it will stop flashing and stay on to show that the battery is...

Страница 10: ...inserted into the jack 2 Line up the dots on the micromotor and contra angle and slide the contra angle straight onto the micromotor until it clicks securely into place The contra angle has a simple...

Страница 11: ...les to avoid injury to fingers Inserting and removing files without holding the file release button may damage the chuck Do not attach the file electrode to files with a shaft diameter of 1 2mm or gre...

Страница 12: ...erly installed Give it a light tug to confirm 5 Check that the file electrode is properly clipped onto the file 6 Check that the contrary electrode is attached to the connector of the handpiece cord 7...

Страница 13: ...the micromotor and the gears Auto Apical Slow Down The file automatically slows down as it approaches the apex so that the region near the apical foramen can be treated with a slow gentle rotation Th...

Страница 14: ...motor connected and press the Mode switch until the speed and memory displays disappear Refer to the operation manual for Canal Measurement Module 2 Detach the micromotor from handpiece cord Select M1...

Страница 15: ...o select M1 M2 or M3 and then perform root canal shaping WARNING When shaping a root canal make sure that the file or contrary electrode does not come into contact with an electric power source such a...

Страница 16: ...root canal length measurement Sound Volume Off Low and High Torque Line 11 settings for auto torque reverse This function may be turned off The motor automatically reverses if the torque exceeds the s...

Страница 17: ...Torque Line Apical Line and Speed settings 5 Press Set switch to enter the settings into the memory Select the item Display will briefly flash on and off Set the memory content Turn sound off Set Sou...

Страница 18: ...rmal or unusual root canal morphology make sure to take an x ray to check the measurement results Stop using the unit immediately if it does not seem to be working properly Refer to the separate manua...

Страница 19: ...is inserted into the root canal Auto Start If the root canal is extremely dry the Auto Start function may not operate NOTE If the Auto Start function does not operate because the root canal is extreme...

Страница 20: ...is applied If the motor overheats do not detach the micromotor from the handpiece cord Even if the motor has cooled down enough to operate it could still be rather hot and excessive force should not...

Страница 21: ...could come out and injure the patient Never press the file release button while the micromotor is running This could cause the button to heat up and burn the patient or the file to come out and injur...

Страница 22: ...anal out with a chemical solution during instrumentation helps stable and consistent file action After root canal preparation clean the root canal out ultrasonically If necessary make minor alteration...

Страница 23: ...measurement display does not appear such as when measuring an extremely dry canal because the motor could start up even if a measurement is not being made 2 Reverse Rotation to release locked file If...

Страница 24: ...speed display stop using the unit immediately and recharge the battery If the plug for the AC adapter does not fit the socket it is the user s responsibility to find a suitable plug adapter When the...

Страница 25: ...charging response or to what is called the memory effect Its normal working condition can be restored in the following way Use the manual mode to run the motor until Lo b low battery appears in the d...

Страница 26: ...e contents in a sterilization pouch for autoclaving Never autoclave micromotor and handpiece cord Remove file before autoclaving contra angle Autoclaving Contra Angle 1 Wipe the contra angle with a pi...

Страница 27: ...ge deform or discolor the module Use extra caution to avoid spilling formalin cresol FC and sodium hypochlorite as they are quite strong Wipe up any chemical spills immediately Some chemicals may leav...

Страница 28: ...m the contra angle 2 Clean the contra angle by wiping it with ethanol Very carefully clean the area where the electrode clips the file and the small screw which retains it in place File electrodes for...

Страница 29: ...Handpiece Module 4 Place the battery cover back on Low Speed Handpiece Module NOTE Be careful not to pinch the battery cord when replacing the cover Dispose of old battery in an environmentally safe w...

Страница 30: ...1 to M2 to M3 etc 3 Check that the Select and Set switches work properly 4 Check that the handpiece cord is properly connected to its jack on Low Speed Handpiece Module and that the contrary electrode...

Страница 31: ...the display is set for root canal preparation or for root canal measurement Display is OK but micromotor will not operate Micromotor handpiece does not operate in manual mode and the overheat indicat...

Страница 32: ...ing a metal prosthesis Are proximal surfaces infected with caries Are there lateral canals or is the tooth fractured Does a broken crown allow the leakage of electric current Is there a lesion at the...

Страница 33: ...o Code No Description 1 File Electrode 2 7504010 Battery 3 7503800 AR Oil 4 7503955 Contra Angle 5 7504060 AC Adapter 120V 6 7504003 7504000 Micromotor TR800 Micromotor TR400 option 7 7503960 Handpiec...

Страница 34: ...the length of the root canal and it can be used as low speed motorized handpiece Main unit Rated Voltage DC 9 6 V with rechargeable battery operation Rated Current max 0 2 A with rechargeable battery...

Страница 35: ...ittent Field Repair It can not be repaired in the field Send to J Morita regional office or local dealer for repair DP ZX Contra Angle Model DP ZX contra angle Free Running Maximum 800 rpm or 400 rpm...

Страница 36: ...EC which includes the requirements for electromagnetic compatibility Serial Number Autoclavable up to 135 C ROOT ZX II conforms with the European Directive 93 42 EEC which includes the requirements fo...

Страница 37: ...t switch Serial Number Connector for the handpiece cord Connector for the cord of the foot switch Line up the marks on the cord and its connector to connect it Input DC 17 V 1 A Direct Current Attenti...

Страница 38: ...ards are disposed as electrical scrap Material must be disposed according to the relevant national legal regulations Consult specialized disposal companies for this purpose Please inquire of the local...

Страница 39: ...ems on page 28 Refer to the parts replacement lists and replace worn parts whenever necessary 1 J Morita Mfg Corp will not be responsible for accidents equipment damage or bodily injury resulting from...